Chapter 4 - General Procedures and Functional Checks
4-3
Section 4-2
General Procedures
4-2-1
Overview
Since some general procedures are used more than others, for convenience, all the procedures
described in this section are listed below:
•
•
•
Log On to the System as ADM section
•
•
Labeling Removable Media section
•
Formatting Removable Media section
•
Verifying Removable Media section
•
Archiving and Loading Presets section
4-2-2
Power ON/Boot-up
4-2-2-1
Connecting Mains Power to the Vivid
i
n/ Vivid q N Ultrasound Unit
DANGER: To avoid the risk of fire, power to the system must be supplied from a properly rated
outlet. It is recommended to use a dedicated power outlet. The power plug should not, under any
circumstances, be altered to a configuration rated less than that specified for the current.
DO NOT use an extension cord or adaptor plug.
Refer to
2-2-5 "Electrical Requirements" on page
DANGER: ALWAYS CONNECT THE UNIT TO A FIXED POWER SOCKET WHICH HAS THE
PROTECTIVE GROUNDING CONNECTOR.
DANGER:
NEVER USE A THREE-TO-TWO PRONG ADAPTER; THIS DEFEATS THE SAFETY
GROUND.
DANGER: ENSURE THAT THE POWER CORD AND PLUG ARE INTACT AND THAT THE POWER
PLUG IS THE PROPER HOSPITAL-GRADE TYPE (WHERE REQUIRED).
CAUTION: The system requires all covers.
Operate this unit only when all board covers and frame panels are securely in place. The covers
are required for safe operation, good system performance and cooling purposes.
NOTE:
When turning on a system from standby mode, it takes a few seconds before it responds. Do not push
the On/off button again during this period. A second push will initiate a full shutdown.
NOTE:
If the unit has been in the OFF condition for an extended period of time, (3 to 5 days or more), the unit
may not boot, or may beep when turned on..
NOTE:
If the system cannot boot up, or it boots up with a PS communication error, refer to the instructions for
BIOS Battery Replacement on page 8-45
NOTICE
Use only power supply cords, cables and plugs provided by or designated by GE Medical Systems.