4D Stress Echo
Exercise/Treadmill
1. Select 4D probe, Exercise application
2. Press
Protocol
3. Select Stress on touch panel
4. Select Template – Exercise 2 X 5 2D/4D
5. Press
Begin/Cont to start the procedure
6. Press
Store to capture baseline images acquired
7. Press
Store to activate continuous capture acquisition
8. Use
< > or to move or skip to next cell
Tips: Press
Protocol Pause – to orient your 4D image
Select
Multi beat or Single beat for 4D in template editor
4D Stress Echo
(4D Stress Echo – cont.)
Pharmacological
1. Select 4D probe, exercise application
2. Press
Protocol
3. Select Stress on touch panel
4. Select Template – Pharm 4 X 5 2D/4D
5. Press
Begin/Cont to start the procedure
6. Press
Store to capture 2D baseline images
Note: pause protocol if need to check image, return to protocol