Back End Processor (BEP) parts replacement
Vivid E80/E90/E95
–
Service Manual
8-101
GC091052
Rev. 3
15. Restore the Patient Archive images, Report Archive,
User-Defined Configuration (Customer Presets), and
Service settings.
Verification after BEP or HDD replacement
Perform the following steps to verify that the product is
functioning as intended after this replacement:
1. Verify that all screws that you removed earlier have been
installed.
2. If finished, connect cables and probes removed earlier.
3. Power up the Ultrasound system to verify that it operates as
intended.
4. Check/set Date, Time, and Time Zone, and set the Preset
Region accordingly.
5. Reconnect to the network.
6. If the Ultrasound system was connected to InSite ExC,
restore the service settings backed up prior to the software
loading. Reboot and verify remote connectivity.
7. Confirm Patient Archive images, Report Archive, User
Defined Configuration (Customer Presets), and Service
settings are restored.
8. If DVD is installed: Format a DVD, and backup Customer
Presets to confirm proper DVD write functionality.
Hard Disk Drive (HDD) replacement
Preparations
WARNING
Before you dispose of the hard drive, make sure you remove
ALL PATIENT DATA from the hard drive, given that the hard
drive is still functional. In some countries, you may be required
to delete all software from the disk before returning the hard
drive to the parts warehouse. Follow your local policies.
Summary of Contents for Vivid E80
Page 4: ...i 2 Vivid E80 E90 E95 Service Manual GC091052 Rev 3 Important precautions Translation policy ...
Page 5: ...Vivid E80 E90 E95 Service Manual i 3 GC091052 Rev 3 ...
Page 6: ...i 4 Vivid E80 E90 E95 Service Manual GC091052 Rev 3 ...
Page 7: ...Vivid E80 E90 E95 Service Manual i 5 GC091052 Rev 3 ...
Page 8: ...i 6 Vivid E80 E90 E95 Service Manual GC091052 Rev 3 ...
Page 9: ...Vivid E80 E90 E95 Service Manual i 7 GC091052 Rev 3 ...
Page 10: ...i 8 Vivid E80 E90 E95 Service Manual GC091052 Rev 3 ...
Page 11: ...Vivid E80 E90 E95 Service Manual i 9 GC091052 Rev 3 ...
Page 12: ...i 10 Vivid E80 E90 E95 Service Manual GC091052 Rev 3 ...
Page 13: ...Vivid E80 E90 E95 Service Manual i 11 GC091052 Rev 3 ...
Page 14: ...i 12 Vivid E80 E90 E95 Service Manual GC091052 Rev 3 ...
Page 15: ...Vivid E80 E90 E95 Service Manual i 13 GC091052 Rev 3 ...
Page 16: ...i 14 Vivid E80 E90 E95 Service Manual GC091052 Rev 3 ...
Page 22: ...i 20 Vivid E80 E90 E95 Service Manual GC091052 Rev 3 THIS PAGE WAS INTENTIONALLY LEFT BLANK ...
Page 30: ...i 28 Vivid E80 E90 E95 Service Manual GC091052 Rev 3 ...
Page 371: ...Vivid E80 E90 E95 Service Manual 10 1 GC091052 Rev 3 Chapter 10 Care and Maintenance ...
Page 381: ......