Functional checks
Vivid E80/E90/E95
–
Service Manual
4-29
GC091052
Rev. 3
Probe/Connectors Checks
NOTE:
Probes can be connected at any time, whether the Ultrasound
system is ON or OFF
CAUTION
Take the following precautions with the probe cables:
•
Keep away from the wheels.
•
Do not bend.
•
Do not cross cables between probes.
Table 4-1:
Probe and Connectors Checks
Step
Task
Expected Result(s)
1.
Press
Probe
on the Operating
Panel.
A list of the connected probes will
pop up on the screen.
2.
If not already selected, use the
trackball to select the desired
probe.
An application menu for the desired
probe is listed on the screen.
3.
•
Trackball to the desired
application.
• Press
Select
to launch the
application.
•
To change application without
changing the current probe,
press
Appl.
on the Operating
Panel.
The selected application starts.
4.
Verify no missing channels.
All channels are functioning.
5.
Verify there's no EMI/RFI or
artifacts specific to the probe.
No EMI/RFI or artifacts.
6.
Test the probe in each active
connector slot.
It will display pictorial data each
time.
7.
Do a leakage test on the probe.
It passes the test.
8.
Repeat this procedure for all
available probes.
Summary of Contents for Vivid E80
Page 4: ...i 2 Vivid E80 E90 E95 Service Manual GC091052 Rev 3 Important precautions Translation policy ...
Page 5: ...Vivid E80 E90 E95 Service Manual i 3 GC091052 Rev 3 ...
Page 6: ...i 4 Vivid E80 E90 E95 Service Manual GC091052 Rev 3 ...
Page 7: ...Vivid E80 E90 E95 Service Manual i 5 GC091052 Rev 3 ...
Page 8: ...i 6 Vivid E80 E90 E95 Service Manual GC091052 Rev 3 ...
Page 9: ...Vivid E80 E90 E95 Service Manual i 7 GC091052 Rev 3 ...
Page 10: ...i 8 Vivid E80 E90 E95 Service Manual GC091052 Rev 3 ...
Page 11: ...Vivid E80 E90 E95 Service Manual i 9 GC091052 Rev 3 ...
Page 12: ...i 10 Vivid E80 E90 E95 Service Manual GC091052 Rev 3 ...
Page 13: ...Vivid E80 E90 E95 Service Manual i 11 GC091052 Rev 3 ...
Page 14: ...i 12 Vivid E80 E90 E95 Service Manual GC091052 Rev 3 ...
Page 15: ...Vivid E80 E90 E95 Service Manual i 13 GC091052 Rev 3 ...
Page 16: ...i 14 Vivid E80 E90 E95 Service Manual GC091052 Rev 3 ...
Page 22: ...i 20 Vivid E80 E90 E95 Service Manual GC091052 Rev 3 THIS PAGE WAS INTENTIONALLY LEFT BLANK ...
Page 30: ...i 28 Vivid E80 E90 E95 Service Manual GC091052 Rev 3 ...
Page 371: ...Vivid E80 E90 E95 Service Manual 10 1 GC091052 Rev 3 Chapter 10 Care and Maintenance ...
Page 381: ......