General Procedures and Functional Checks
4-12
Vivid E80/E90/E95
–
Service Manual
GC091052
Rev. 3
• Use
the
Depth
control to adjust the range to be imaged.
• Use
the
Frequency
control (move to higher frequencies) or
the
Frame rate
control (move to lower frame rate) to
increase resolution in image.
• Use
the
Frequency
control (move to lower frequency) to
increase penetration.
• Use
the
Reject
control to reduce noise in the image.
• Use
the
DDP
control to optimize imaging in the blood flow
regions and make a cleaner, less noisy image.
• Use
UD Clarity
(Cardiac) or
UD Speckle reduce
(non-cardiac) to reduce image speckle. Extra care must be
taken to select the optimal Speckle reduction level, as too
much filtering of speckle can mask or obscure desired
image detail.
• Adjust
Octave
to toggle between Fundamental and
Harmonic mode.
• Press
Color maps
and select a gray map from the menu on
screen.
•
If using a 4D probe:
• Adjust
the
Quick Rotate
control on the Touch panel or
press
Angle
on the Control panel to rotate the scan
plane to predefined angles.
• Adjust
the
Rotate
rotary of the Touch panel to fine tune
the angle adjustment.
A scan plane indicator is displayed showing the angle
position of the scan plane.
Summary of Contents for Vivid E80
Page 4: ...i 2 Vivid E80 E90 E95 Service Manual GC091052 Rev 3 Important precautions Translation policy ...
Page 5: ...Vivid E80 E90 E95 Service Manual i 3 GC091052 Rev 3 ...
Page 6: ...i 4 Vivid E80 E90 E95 Service Manual GC091052 Rev 3 ...
Page 7: ...Vivid E80 E90 E95 Service Manual i 5 GC091052 Rev 3 ...
Page 8: ...i 6 Vivid E80 E90 E95 Service Manual GC091052 Rev 3 ...
Page 9: ...Vivid E80 E90 E95 Service Manual i 7 GC091052 Rev 3 ...
Page 10: ...i 8 Vivid E80 E90 E95 Service Manual GC091052 Rev 3 ...
Page 11: ...Vivid E80 E90 E95 Service Manual i 9 GC091052 Rev 3 ...
Page 12: ...i 10 Vivid E80 E90 E95 Service Manual GC091052 Rev 3 ...
Page 13: ...Vivid E80 E90 E95 Service Manual i 11 GC091052 Rev 3 ...
Page 14: ...i 12 Vivid E80 E90 E95 Service Manual GC091052 Rev 3 ...
Page 15: ...Vivid E80 E90 E95 Service Manual i 13 GC091052 Rev 3 ...
Page 16: ...i 14 Vivid E80 E90 E95 Service Manual GC091052 Rev 3 ...
Page 22: ...i 20 Vivid E80 E90 E95 Service Manual GC091052 Rev 3 THIS PAGE WAS INTENTIONALLY LEFT BLANK ...
Page 30: ...i 28 Vivid E80 E90 E95 Service Manual GC091052 Rev 3 ...
Page 371: ...Vivid E80 E90 E95 Service Manual 10 1 GC091052 Rev 3 Chapter 10 Care and Maintenance ...
Page 381: ......