TruSat User’s Guide
4–2
Specifications
Specifications are nominal and are subject to change without notice.
Measurement
General
Pulse oximetry sensors: sensors only
Method: red and infrared light absorption
Red LED wavelength range: 650 to 670 nm
Infrared (IR) LED wavelength range: 930 to 950 nm
Average power:
≤
1 mW
SpO
2
Calibrated for functional oxygen saturation
Calibration range: 70 to 100%
Measurement and display range: 1 to 100%
Display resolution: 1%
First reading, full accuracy:
≤
10 seconds
Accuracy, Arms (root mean square of paired values; previously represented by
± 1 standard deviation)
70 to 100% ± 2 digits (without motion)
70 to 100% ± 3 digits (during clinical motion)
2
70 to 100% ± 2 digits (during clinical low perfusion)
Below 70% unspecified
SpO
2
measurement accuracy is based on deep hypoxia studies using sensors
on volunteer subjects. Arterial blood samples were analyzed simultaneously on multiple
CO-oximeters.
NOTE: Accuracy may vary for some sensors; always check the instructions for the
sensor.
Pulse rate
Measurement and display range: 30 to 250 beats per minute (bpm)
Display resolution: 1 bpm
First reading, full accuracy:
≤
15 seconds
Accuracy
30 to 250 bpm: ± 2 digits or ± 2%, whichever is greater, (without motion)
30 to 250 bpm: ± 5 digits (during motion)
30 to 250 bpm: ± 3 digits (during low perfusion)
PIr pulsatile value
Measurement and display range: 0.01 to 9.99
Display resolution: 0.01 PIr
2
Applicability: Adult/Pediatric and AllFit sensors.