![GE Telligence Patient-Staff Communications System Operation Manual Download Page 113](http://html.mh-extra.com/html/ge/telligence-patient-staff-communications-system/telligence-patient-staff-communications-system_operation-manual_132154113.webp)
Chapter 5
Staff/duty stations
Telligence Patient-Staff Communication System Operation Manual
107
smoke detector that activates, the call station places a high-priority
room alarm nurse call. When the smoke detector alarm ends, the
high-priority nurse call clears.
1. Press the Call button (or pull the optional cord) on the call
station to send a staff normal call to staff consoles and
annunciators.
Pressing the Call button (or pulling the cord) causes the
following to happen:
• lights by the station‘s Call button and by any linked Call
buttons turn on to confirm nurse call placement.
• The call assurance tone sounds at any associated patient
or staff/duty station.
• Associated corridor and zone lights show a Input (shown
without pull cord) steady white.
• Tones sound at staff consoles and unit annunciators, and at
stations with staff follow active.
• A staff normal nurse call row appears on the Call screen on
staff consoles and unit annunciators.
Note:
A staff normal call is cleared from the system when a staff
member answers the call at a staff console or annunciator.
2. To cancel a call from the model HC-PB2-CALLIN call station (that
was not already cleared by a staff member answering the call),
press Cancel on the originating call station, or a linked Cancel
button on any associated patient or staff room station.
• Call button lights extinguish
• Dome lights extinguish.
• Tones stop sounding at staff consoles and annunciators,
and at stations where staff follow is active.
• The nurse call row disappears from staff consoles and
annunciators.
If the nurse call had a reminder set for it, the following happens
when the Cancel button is pressed:
• Dome light display of the staff level reminder ceases.
• The reminder row disappears from staff console or
annunciator Reminders screens.
Call stations with relay
The model HC-PB2-CALLR call station is a generic nurse call station.
This station may be equipped with a pull cord, may be a standalone
station, or may be associated with a patient or staff/duty station.
The call station may be installed in patient rooms or other rooms
where patients are temporarily located, such as exam or recovery
rooms. It may also be located in staff rooms. The station, which