6. All metallic pipes (including water, medical gas, sprinklers, etc.) entering the RF Shield,
excluding the Cryogenic Vent and floor drains, must be located within 56 inches (1422 mm)
of the RF Common Ground Stud
7. The illustration below shows a typical ground layout
Illustration 3-22: Typical Magnet Room Grounding
SIGNA Voyager Pre-Installation
Direction 5680008–1EN, Revision 2
96
9 Magnet Room Electrical and Grounding Requirements