2 MR Suite Minimum Room Size Requirement
CAUTION
Procedure Failure
The minimum service area shown must be kept clear of permanent or installed
cabinetry, mill work, shelving, coil storage fixtures, furniture, etc. Ceiling service
area should be kept clear of overhead items, including soffits, HVAC, plumbing
compnent and brackets
Permanent or installed objects in this area may prevent or delay magnet service
or operation.
Room dimensions shown in the table below lists the minimum finished room space
requirements to properly and safely operate and service the MR system. The items listed below
are not included in the minimum area dimensions:
1. Building code requirements (e.g., exit routes, door placement, local and national electrical
codes, etc.)
NOTE:
The customer must provide Equipment and Magnet Room evacuation routes to
comply with facility emergency procedures.
2. System requirements, including cable run locations, cryogen venting, patient observation
requirements, and penetration panel placements (e.g., the Equipment room and Magnet
room must share a common wall to allow penetration panel installation)
3. Penetration panel closet and all associated areas (must be outside the minimum finished
room dimensions)
4. Non-GEHC equipment options (such as additional AC or water cooling equipment in the
Equipment room)
5. Accessory storage. Refer to the
Customer Site Storage Requirements
manual (document
number 5182674) or contact the GE Healthcare Project Manager of Installation (PMI) for
any additional accessory storage requirements
Table 2-1: System Minimum Room Inside Clear Space Dimensions
Equipment Room
Magnet Room
Control Room
Total System
Area
ft
2
(m
2
)
W x D
in. (mm)
Area ft
2
(m
2
) Ceiling Height
In. (mm)
W x D
In. (mm)
Area
ft
2
(m
2
)
Ceiling
Height
in. (mm)
W x D
In. (mm)
Area
ft
2
(m
2
)
107.7 x 70.9
(2735 x 1800)
52.9
(4.92)
114
(2896)
145.8 x
228.8
(3704 x
5812)
See Note
1,2,3
201.4
(18.7)
98.5
(2500)
See Note 4
83.9 x 59.8
(2130 x
1520)
34.8
(3.24)
293.2
(27.24)
SIGNA Voyager Pre-Installation
Direction 5680008–1EN, Revision 2
Chapter 2 General System Level
23