MINIMUM MAGNET CEILING HEIGHT (TOP VIEW)
Shaded area within solid lines indicates floor to ceiling minimum height of 2500 mm [98.42in]. If ceiling is between
2500 mm [98.42in] and 2667 mm [105in], flexible main lead extension for low ceiling height is required for
ramping the magnet. Contact the PMI for preparation.
1372
[54 in]
1372
[54 in]
2769
[109 in]
MAGNET
GEOMETRIC
ISOCENTER
FRONT OF MAGNET
SCALE 1:50
59
[2.32 in]
824
[32.44 in
]
942
[37.1 in]
1226
[48.27 in]
1107.2
[43.59 in
]
59
[2.32 in]
59
[2.32 in]
Mesh shield
frame area
Finished wall
6140
156
[
241.74 in
6.15 in
]
Finished
wall
RF Shield
PENETRATION PANELS WALL OPENINGS
Integrated System Cabinet and Integrated Cooling Cabinet must be covered on both sides for safety.
Equipment
room
Magnet room
INTEGRATED SYSTEM CABINET
Front View
INTEGRATED COOLING CABINET
Front View
Equipment room side
EQUIPMENT ROOM FLOOR
INTEGRATED SYSTEM CABINET
Top View
INTEGRATED COOLING CABINET
Top
View
PENETRATION PANEL CLOSET
An enclosure (i.e. closet) must be provided to restrict access to the PEN panels and for storage of excess interconnections.
The PEN closet must have a mechanical locking mechanism to restrict access to the PEN panels
The PEN closet must maintain the minimum service area outside the 200 Gauss in the magnet room.
PEN closet must allow free air exchange of
between the Magnet room and PEN closet for MR
system blowers. Airflow may be achieved through door louvers or other openings in the PEN closet that meet all other
PEN closet requirements
A:
300 mm [11.81in] side space for excess cable sorage. Left side space and right side space can be swapped.
NOTE:All RF mesh Assy Shipping brackets should be kept on frame until after installed and RF test is completed.
RF
shield
ICC/ISC
minimum closet
inner dimensions
outline
A
570
[22.4 in]
570
[22.4 in]
430
[16.9 in]
430
[16.9 in]
122
[4.8 in]
2600
[102.4 in]
400
[15.7 in]
200
[7.9 in]
50
[2 in]
Date
Rev
/24
EN-MRI-TYP-SIGNA_PIONEER-WEB.DWG
A
23/Jul/2018
SIGNA PIONEER
Typical
RF shielding
07