CABLE TRAYS REQUIREMENTS IN MAGNET ROOM
2
7
4
3
8
7
213 mm [8.39 in]
Min bend radius
1 - Ceiling
2 - Finished Floor
3 - Magnet isocenter
4 - Minimum cable tray height required at back of Magnet: 2436 mm [95.9 in].
Tray height may be lower at other points to avoid obstructions.
5 - Maximum height from floor to top of tray (anywhere in Magnet room): 3251 mm [128 in].
6 - Minimum distance from top of cable tray to ceiling: 254 mm [10 in].
Minimum distance on either side of an obstruction: 254 mm [10 in], minimum distance from top of cable tray to
Obstruction: 178 mm [7 in].
9 - Minimum distance between trays: 12 mm [0.5 in].
10 - Non-ferrous cable support
center.
Cable Tray Requirements (Side-By-Side)
1
6
5
10
8
9
11
CABLE TRAYS IN EQUIPMENT ROOM
CABLE TRAY DETAIL
NOT TO SCALE
Min. 300
[11.8 in]
450
[18 in]
RF Shield
ISC
Date
Rev
/24
EN-MRI-TYP-SIGNA_PIONEER-WEB.DWG
A
23/Jul/2018
SIGNA PIONEER
Typical
E4 - Electrical Details