Applications (Apps) – General Information
54
SEER™ 1000
2067634-077 Revision B
NOTE
If the recorder has started automatically or has been manually
started using the
Event Button
, a protection key is automatically gen-
erated by the recorder. In this case, wireless access is no longer pos-
sible until the data has been downloaded and deleted.
Should a recorder become locked because a technician outside the con-
trol of the administrator has established a connection with an empty re-
corder, the recorder can be reset using the PC App as explained in section
“Resetting the Recorder with the Microsoft Windows App” on page 90.
NOTE
The Apple iOS version of the application does not allow the recorder
to be reset.
CardioDay users may also use CardioDay rather than an app to download
patient demographic data to a recorder before starting a recording. If the
CardioDay administrator also has the same password as the app admin-
istrator, then the ECG technicians may use their apps to communicate
with the recorders.
Basic Information on Wireless Bluetooth Connection
NOTICE
NO OPTIMUM PERFORMANCE – If you are using any NON-GE SUP-
PLIED Bluetooth wireless equipment, optimum performance cannot
be assured.
Only use the Bluetooth wireless equipment supplied BY GE, as the Blue-
tooth wireless technology has been tested and validated using this
equipment.
Bear in mind, EVEN WHEN USING GE SUPPLIED BLUETOOTH WIRELESS
EQUIPMENT that the complete data transfer cannot be guaranteed
under all circumstances when using Bluetooth wireless technology
and that range loss can occur if Bluetooth wireless technology and
other RF devices (e.g., WLAN) are used near each other. Also, other
devices could interfere with the equipment, even if the other devices
comply with CISPR emission requirements.