Introduction
20
SEER™ 1000
2067634-077 Revision B
Symbol
Description
Catalog or Orderable Part Number
Indicates the manufacturer's catalog or part number.
Serial Number
Indicates the manufacturer's serial number.
Manufacturer Name and Address
Indicates the name and address for the manufacturer of
this device.
Date of Manufacture (Year)
Indicates the original manufacture date for this device.
Consult Instructions for Use
Consult the operating instructions.
Type CF Applied Part
Identifies a type CF applied part on medical equipment
that complies with IEC 60601–1.
This device meets the requirements for protection against
electric shock for an earth-free (floating) applied part (one
intended for contact with patients) for cardiac application.
IP43
IP Code (Ingress Protection Rating)
Classifies and rates the degree of protection provided
against the intrusion of solid objects (such as body parts
like hands and fingers, dust, accidental contact), and fluids.
The first numeral (4) represents the degree of protection
against the ingress of solid objects. In this case, objects
> 1 mm (0.04 inches).
The second numeral (3) represents the degree of protec-
tion against the ingress of liquids. In this case, protected
against spraying water.
NOTICE
INGRESS PROTECTION RESTRICTED TO THE FOLLOW-
ING CONDITIONS – IP43 applies only when the bat-
tery door is closed and latched AND if the ECG cable
is properly connected to the recorder.
Before starting a recording, make sure that the bat-
tery door is closed and latched AND that the ECG ca-
ble is properly connected to the recorder.