Preparing the Recording
40
SEER™ 1000
2067634-077 Revision B
WARNING:
SHOCK HAZARD — Touching the conductive elements can-
cels the protection provided by isolated signal input.
Ensure that conductive parts of the electrodes or leadwires
do not come in contact with other conductive parts.
4.
Look at the
Hookup Guidance
LED
for indication of lead problems.
NOTE:
Use only electrodes and contact agents recommended by
GE Healthcare.
Placing the Electrodes
This section describes the placement of the electrodes for 2-channel and
3-channel Holter ECG.
CAUTION
PROPER LEAD WIRE CONNECTION — Improper connection will cause
inaccuracies in the ECG.
Use stress loops to reduce the movement of the electrodes caused by
tension from the leadwires.
Secure the electrodes with tape to prevent movement and maintain
electrical contact.
Apply the tape loosely enough so that no gel squeezes onto the adhe-
sive surfaces of the electrodes.