PACE Series Calibration Manual

[EN] English

1

K0450 Revision B

1

Introduction

The PACE controller incorporates a calibration facility; for the PACE to stay accurate, a 
calibration check should be carried out at chosen intervals. If the accuracy of the PACE is not 
within the permissible deviation, carry out a calibration adjustment.

2

Calibration Status

Using the 

Measured Pressure/Instrument Status

 menu, the calibration status of the 

calibrator can be displayed on the front panel screen. The status menu includes 

Calibration 

History

 which gives a list of dates of the stored calibration corrections.

Note:

The Date and Time must be set correctly using the 

Measured Pressure

/Global Set-up/

Calibration

 menu.

3

Calibration Equipment

The original GE Calibration Certificate shows the measurement uncertainty of the original 
calibration standard. To preserve uncertainty of the PACE calibration, checks and 
adjustments must be performed using a calibrator uncertainty of less than or equal to the 
original calibration standard.

4

Preliminary Operations

Review and become familiar with the whole procedure before beginning a calibration 
process. Allow at least one hour for the PACE to thermally stabilize in a thermally stable 
environment after switching on and before calibration.

Before starting a calibration procedure:

Carry out a leak test as detailed in PACE user manual K0443.

5

Notes on Calibration

The pressure standard output port and the reference 
level must be at the correct level or use height-
corrected applied pressure.
To prevent applied calibration pressure “back feed” in 
PACE controllers, fit blanking plugs to both positive and 
negative supply ports on the manifold.
Set the PACE units of pressure to one of the required units for calibration.

Reference 

Level

Summary of Contents for PACE5000

Page 1: ...GE Digital Solutions Pressure Automated Calibration Equipment Calibration manual K0450 Revision B PACE5000 PACE6000 PACE1000...

Page 2: ...ons are subject to change without notice GE is a registered trademark of General Electric Company Other company or product names mentioned in this document may be trademarks or registered trademarks o...

Page 3: ...ge Use suitably qualified calibration technicians and good engineering practice for all procedures in this publication Pressure Do not apply pressures greater than the maximum working pressure to the...

Page 4: ...d airspeed i e That is IEC International Electrotechnical Commission IEEE 488 Institute of Electrical and Electronic Engineers standard 488 data in Inch kg Kilogram kts kn knot LCD Liquid crystal disp...

Page 5: ...ets the requirements of all relevant European safety directives The equipment carries the CE mark This symbol on the equipment indicates that the user should read the user manual This symbol on the eq...

Page 6: ...RE APPLYING PRESSURE EXAMINE ALL FITTINGS AND EQUIPMENT FOR DAMAGE REPLACE ALL DAMAGED FITTINGS AND EQUIPMENT DO NOT USE ANY DAMAGED FITTINGS AND EQUIPMENT DO NOT EXCEED THE MAXIMUM WORKING PRESSURE O...

Page 7: ...r the list of selectable pressure units Pressure units Factor hPa Pressure units Factor hPa mbar 1 0 cmH2O 20 C 0 978903642 bar 1000 0 mH2O 20 C 97 8903642 Pa N m2 0 01 kg m2 0 0980665 hPa 1 0 kg cm2...

Page 8: ...es on Calibration 1 5 1 PACE Controller Connection 2 Gauge Reference 2 5 2 PACE Indicator Connection 3 Gauge Reference 3 6 Calibration Check All Ranges 4 6 1 Procedure 4 7 Calibration Adjustment 5 App...

Page 9: ...rement uncertainty of the original calibration standard To preserve uncertainty of the PACE calibration checks and adjustments must be performed using a calibrator uncertainty of less than or equal to...

Page 10: ...open to atmosphere Turn the valve to the closed position Gauge Reference If the pressure standard has a reference connection then connect this to the PACE reference port on the module manifold Otherw...

Page 11: ...essure standard has a reference connection then connect this to the indicator reference port Otherwise the indicator reference port should be open to atmosphere Note Indicator reference port open to a...

Page 12: ...e to the first pressure value and wait until this pressure as displayed on PACE is stable to less than 10ppm 0 001 Display filtering may be required 4 Compare the pressure value on the calibration sta...

Page 13: ...e to be set on the pressure standard and to press OK when the applied pressure is stable Use the numeric keys to enter the precise applied pressure Note The display also shows throughout this procedur...

Page 14: ...K0450 Revision B 6 EN English Appendix A PACE Menus and Screens A 1 PACE Controller Menus Calibration procedure next page...

Page 15: ...PACE Series Calibration Manual EN English 7 K0450 Revision B A 2 PACE Controller Screens...

Page 16: ...K0450 Revision B 8 EN English A 3 PACE Indicator Menus Calibration procedure next page...

Page 17: ...PACE Series Calibration Manual EN English 9 K0450 Revision B A 4 PACE Indicator Screens...

Page 18: ......

Page 19: ......

Page 20: ...www gemeasurement com...

Reviews: