PACE Series Calibration Manual
[EN] English
3
K0450 Revision B
5.2 PACE Indicator Connection
Connect the indicator for each pressure range as follows:
Connect the output of the pressure standard to the indicator port.
Gauge Reference
If the pressure standard has a reference connection, then connect this to the indicator
reference port. Otherwise the indicator reference port should be open to atmosphere.
Note:
Indicator reference port open to atmosphere or to reference port of the pressure
standard.
Note:
Optimum performance is achieved by connecting the PACE reference port to the
pressure standard.
SENSOR
PORT
PACE Indicator
Rear view of
instrument
Pressure Standard
REFERENCE
PORT
Summary of Contents for PACE5000