GE Analytical Instruments ©2006
9-6
DLM 14291 Rev. A
The NOAnalysis Bag program can be used for collecting data during analysis of the
chamber sampling (consult the NOAnalysis manual for instructions). Alternatively,
the NOA’s front panel display can be used for viewing the concentration of NO in
the samples.
NOA Setup
For the off-line NO measurements, the NOA can be in either the Exhalation Mode
or the Nitric Oxide mode with the Units set to Concentration. Sensitivity should be
set to Auto and the Com Port Interval set to 1/2 or 1/4 second. See Chapter 5
(INSTALLATION AND SETUP: GAS-PHASE MEASUREMENTS) for configuring the NOA.
The NOA should be calibrated (Chapter 6 – CALIBRATION) prior to measurement.