Figure 8- 2: Diagram of Deadspace Discard Bag Collection Kit.
•
Screw the pressure gauge into the bag filler body (Teflon tape is not required).
•
Inspect the pressure gauge. If the needle does not read 0 cm H
2
O, unscrew the
clear plastic cover of the gauge and adjust the screw at the base of the gauge
with a screwdriver until the needle points to 0 cm H
2
O. Replace the clear
plastic cover of the gauge.