background image

 

 

 

 

GE OSMONICS 

MILLENIUM™ REVERSE OSMOSIS UNIT 

 

Operation 

and Maintenance Manual

 

 
 
 

 

 

 

 

 

 

SUPERCEDES P/N 14693 REV. D 

Summary of Contents for MILLENIUM OSMONICS MILLENIUMTM REVERSE OSMOSIS...

Page 1: ...GE OSMONICS MILLENIUM REVERSE OSMOSIS UNIT Operation and Maintenance Manual SUPERCEDES P N 14693 REV D...

Page 2: ......

Page 3: ...3 1 10 4 1 10 5 1 10 6 Specifications Inlet Water Requirements Flow Rates Electrical Requirements Dimensions Membrane Performance Characteristics Environmental Requirements 1 12 1 11 Disposal 1 13 1 1...

Page 4: ...2 BIOSAN Pre Cleaning Preparation 4 14 4 3 2 1 BIOSAN Cleaning Procedure 4 14 4 3 2 2 BIOSAN Recirculation and Dwell Time 4 15 4 3 2 3 BIOSAN Post Cleaning Rinse Procedure 4 16 4 4 Millenium Disinfec...

Page 5: ...ff 5 6 5 5 Low Pump Pressure 5 7 5 6 Pressure Gauge Vibrates Erratically 5 7 5 7 Excess Product Pressure too high 5 7 5 8 Product Flow Too Low 5 8 5 9 Product Flow Too High 5 8 5 10 Leaks 5 9 5 11 Low...

Page 6: ...14693e D D 31 March 2003 This page intentionally left blank...

Page 7: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 1 1 General Information Millenium RO System CHAPTER ONE GENERAL INFORMATION...

Page 8: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 1 2 General Information This page intentionally left blank...

Page 9: ...sary of relevant terms This safety summary does not contain all of the safety statements in the manual Other safety statements are included within the manual text and are enhanced and defined as follo...

Page 10: ...in an explosive environment or near flammable anesthetics 1 3 Applications The Millenium Reverse Osmosis RO System is designed to purify water by forcing water through a semi permeable RO membrane Wa...

Page 11: ...mum capacity of 2 gallons per RO per minute of continuous flow for as long as and as often as the RO is operating TEMPERATURE The Millenium RO is normally connected to a cold water supply The performa...

Page 12: ...ections in this system WARNING Ordinary pipe dope must be avoided since it may leach objectionable and potentially dangerous impurities into the water 1 6 Theory of Operation The Millenium Portable Re...

Page 13: ...mounted outside the RO above the product pressure gauge The pressure is factory set at 190 PSI Reverse Osmosis Membrane Assemblies ROMAs Two are included in the Millenium RO A ROMA is an assembly that...

Page 14: ...oduct side of the RO membranes CAUTION Back flow pressure into the product line can cause damage to the membranes Total Dissolved Solids TDS Sensor Samples the product water quality via the sensor pro...

Page 15: ...reads the product pressure from 0 to 100 PSI Product pressure will read zero when the product line is not connected to a dialysis unit and the product hose is connected to the product fitting and the...

Page 16: ...performance under optimum conditions Product Water Total Dissolved Solids TDS The Product Water Total Dissolved Solids TDS digital display indicates product water purity in mg L ppm It is temperature...

Page 17: ...change head styles WARNING Carbon is not regenerated and must be replaced when exhausted WARNING A suitable carbon filter should always be present as pretreatment for the RO unit in order to prevent c...

Page 18: ...machine pH operating 4 0 11 0 TF Membranes pH w chloramines present operating 4 0 8 5 TF Membranes pH cleaning 2 0 11 0 TF Membranes Pressure dynamic 30 PSI 100 PSI 1 10 2 Flow Rates Product Flow base...

Page 19: ...ting Pressure 190 PSI 10 PSI Product Back pressure 0 PSI 1 10 6 Environmental Requirement Minimum Maximum Ambient Temperature 4 C 39 F 32 C 90 F Storage Temperature 2 C 36 F 32 C 90 F Altitude N A 10...

Page 20: ...chnician or manufacturer for help 1 13 Service Assistance If service assistance is required please take the following steps Consult the Troubleshooting section of this manual Chapter Five If the probl...

Page 21: ...quest Technical Service A GE Osmonics Technical Service representative will provide instructions and a return authorization number which needs to be clearly written on the outside of the box used to s...

Page 22: ...of Medical Instrumentation CA Cellulose Acetate C Celsius CSA Canadian Standards Association cc Cubic Centimeters cf Cubic Foot feet cfu ml Colony Forming Units per milliliter cm Centimeters DI Deioni...

Page 23: ...itation Foundation PA Polyamide P N Part Number ppm Parts Per Million psi Pounds Per Square Inch PVC Polyvinyl Chloride QD Quick Disconnect RO Reverse Osmosis ROMA Reverse Osmosis Membrane Assembly TB...

Page 24: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 1 18 General Information NOTES...

Page 25: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 2 1 Installation Guidelines and Instructions Millenium RO System CHAPTER 2 INSTALLATION GUIDELINES AND INITIAL SYSTEM START UP...

Page 26: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 2 2 Installation Guidelines and Instructions This page intentionally left blank...

Page 27: ...tem the carbon tank must be completely rinsed of all carbon fines and ash before connecting to the RO unit Otherwise irreparable damage to the TF membrane will result refer to Carbon Manual WARNING Th...

Page 28: ...alve to the flush position 7 Turn the inlet water valve switch on Water will start to fill the RO Allow the RO to rinse to drain for five minutes after water begins to flow from the drain line The pum...

Page 29: ...er setting at start up WARNING The TDS alarm limit should not be moved from this setting unless instructed by a physician 12 After a minimum of 3 hours run time disinfect and rinse the RO unit with an...

Page 30: ...e performance verified by the appropriate testing 2 Connect the RO unit to the required utilities feed water drain power 3 Operate the RO unit with product and waste flows to drain for at least 15 min...

Page 31: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 2 7 Installation Guidelines and Instructions This page intentionally left blank...

Page 32: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 3 1 1 System Operation Millenium RO System CHAPTER 3 SYSTEM OPERATION...

Page 33: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 3 2 2 System Operation This page intentionally left blank...

Page 34: ...monitor should indicate 95 rejection within 2 minutes of start up It is normal for rejection to drop off until membranes have rinsed up to optimum quality 7 The TDS monitor should stabilize within 2...

Page 35: ...hose When the pressure is released turn the switch off NOTE If the Millenium RO is to be stored for more than 24 hours RO not being operated the RO should be stored in an approved disinfectant or sto...

Page 36: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 3 5 5 System Operation NOTES...

Page 37: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 3 6 6 System Operation This page intentionally left blank...

Page 38: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 4 1 Routine Maintenance Millenium RO System CHAPTER FOUR ROUTINE MAINTENANCE...

Page 39: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 4 2 Routine Maintenance This page intentionally left blank...

Page 40: ...the feed water in the pretreatment log 2 Observe and record salt level in the brine tank for systems with water softener pretreatment 3 Observe and record the Chlorine Chloramine concentrations in th...

Page 41: ...e pre filtration CAUTION On occasion the pressure differential between the two pre filter gauges will not alter significantly however the pre filter will be soiled If the core of the pre filter is dir...

Page 42: ...nce 1 An AAMI Water Quality Analysis of the product water should be conducted at least yearly Refer to the Water Quality Analysis Procedure in Chapter two 2 Check the monitoring for accuracy 3 Check t...

Page 43: ...r to performing this maintenance on any threaded fitting release the pressure on the piping tubing system i e turn off the pump open a drain port etc 2 Remove the connecting tubing 3 Re tape the pipe...

Page 44: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 4 7 Routine Maintenance Tubing connections...

Page 45: ...side to dry on a clean paper towel 7 Clean the inside of the fitting with a wet cotton swab or paper towel to remove any dirt or debris Be sure not to leave any cotton fibers from the swab or towel in...

Page 46: ...e flow completely Measure Product water flow again and compare against previous results If there is a noticeable increase in flow the check valve needs replaced 3 If none of the above occurs check val...

Page 47: ...e RO membrane performance at its best Cleaning removes the silt organic and mineral deposits that collect on the RO membrane surface which if left untreated encourage bacterial growth and will eventua...

Page 48: ...tory protection nor special ventilation are required Materials Required Liqui Clean 403 Low pH liquid GE Osmonics Part No 1228653 1 Gal Liqui Clean 211 High pH liquid GE Osmonics Part No 1228791 1 Gal...

Page 49: ...e 2 Push and hold the disinfect switch to operate the RO pump for three minutes CAUTION Some pump noise chattering is normal in the flush position Excessive pump noise chattering can be damaging to th...

Page 50: ...tion Let water flow to the drain for one minute 6 Rotate the operate flush valve to the operate position 7 Set the RO pump power switch in the ON position Rinse for 20 minutes with the Operate Flush v...

Page 51: ...ce 11 Turn the RO pump and inlet water valve switches OFF Disconnect the product rinse line and drain and store it 12 Reconnect the product line to its original connection 13 Remove the Do Not Use Con...

Page 52: ...r hose with male QD fitting on one end open on the other DO NOT USE warning labels not provided by Osmonics cleaning disinfect log sheets most current log sheets for water quality comparison 4 3 2 1 B...

Page 53: ...10 seconds If not press the disinfect switch to turn RO pump off and verify that the disinfect line is connected properly CAUTION DO NOT ALLOW PUMP TO RUN DRY Damage will occur if pump is run dry CAUT...

Page 54: ...sition Rinse for 20 minutes with the Operate Flush valve in the Operate position until the rejection and TDS monitor displays normalize e g usual rejection and usual TDS reading Refer to your operatin...

Page 55: ...are not recommended disinfectants for the Millenium 4 4 1 Formaldehyde Disinfection Preparation CAUTION New replacement thin film membranes must be rinsed to drain with normal pressures for a minimum...

Page 56: ...ical setting in a 37 solution Check the labeling on the packaging to make certain this is the case If the solution is other than 37 the dilution will require adjustment The end solution should not exc...

Page 57: ...d of 4 4 4 1 3 Formaldehyde Post Disinfection Rinse Procedure 1 Connect the product rinse line input water and waste lines to the RO 2 Ensure that the product rinse and waste lines are in the drain 3...

Page 58: ...ning signs from the RO only after a negative test confirms the absence of formaldehyde 12 The Millenium RO System is now ready for use NOTE Success of the disinfection procedure may be ascertained by...

Page 59: ...nimum of three 3 hours before first exposure to Renalin No mandatory pre rinse of the membranes is required for future disinfections WARNING Follow all Material Safety Data Sheet MSDS precautions and...

Page 60: ...ection procedure will raise the temperature of the solution The operator must be prepared to add ice to the solution or discontinue the procedure when the water temperature approaches 77 F 25 C 1 Read...

Page 61: ...r to be drawn up into the RO unit as this will damage the pump 6 Momentarily turn the inlet water valve switch on approximately 5 seconds This allows the water to rinse the disinfectant from the disin...

Page 62: ...minute Turn the RO pump switch on for 2 3 minutes This will aid in the rinsing out of any potentially trapped disinfectant in the Millenium RO 8 Continue to rinse for an additional 5 minutes 9 Check t...

Page 63: ...oduct water is 200 colony forming units A bacterial count above the maximum level may be harmful to the hemodialysis patient OTHER Consult your lab to find out what type of sterile container to use an...

Page 64: ...oduct RO water to flow to the drain or allow product water to flow for 30 seconds 6 Maintaining aseptic technique hold the product line over the sterile container with the product water flowing 7 Fill...

Page 65: ...WARNING Always wear rubber gloves an apron and a facemask If the room has ventilation use it If not wear a face respirator with an organic cartridge to remove formaldehyde fumes DO NOT BREATHE FUMES...

Page 66: ...ated in the RO 4 Disconnect the disinfect line waste line feed water line power and product line 5 Coil the hoses and power cord and cover the RO with a large plastic bag to protect the RO from dust C...

Page 67: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 4 30 Routine Maintenance NOTES...

Page 68: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 5 1 Troubleshooting Millenium RO System CHAPTER FIVE TROUBLESHOOTING...

Page 69: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 5 2 Troubleshooting This page intentionally left blank...

Page 70: ...ot 5 6 5 4 Water Produced While RO is Off 5 6 5 5 Low Pump Pressure 5 7 5 6 Pressure Gauge Vibrates Erratically 5 7 5 7 Excess Product Pressure too high 5 7 5 8 Product Flow Too Low 5 8 5 9 Product Fl...

Page 71: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 5 4 Troubleshooting This page intentionally left blank...

Page 72: ...meter c Replace switch 5 Power switch inoperative a Assure inlet water valve is open b Check switch with voltmeter c Replace switch 6 Pump motor inoperative a Verify that voltage supply is correct and...

Page 73: ...a d See 5 5 e Replace pump 2 Fan obstructed or inoperative a Verify no debris is obstructing fan vent b Verify fan operation 3 Excessive current draw a Verify that flow rates and pressures are within...

Page 74: ...ter temperature and flow rate b Check quality of product water B Faulty inlet pressure regulator a Replace regulator C Pump adjusted improperly a Adjust pump at pump head and regulator D Pump worn out...

Page 75: ...truction defective check valve in the product flow path may cause reduced flow 1 Pump pressure too low a See 5 5 2 Low feed water temperature a Verify product production with Tech Note 121 Temperature...

Page 76: ...or blockage or kink in waste line c Quick disconnect not connected completely B High pH and chlorines in feed water a Check feedwater for higher than 8 5 and the presence of chloramines b Check pH red...

Page 77: ...ity tests normal investigate the monitoring circuits 1 Waste flow too low a Disinfectant or cleaner not rinsed from RO b Check waste flow control c Check for obstruction or kink in waste line 2 Pump p...

Page 78: ...quality tests normal investigate the monitoring circuits 1 Waste flow too low a Measure and compare to original data Correct as necessary 2 Waste pressure too low a Adjust pump settings 3 Product flo...

Page 79: ...leak on disinfect draw line a Check that quick disconnect fitting is connected securely b Immerse tube in solution c Check tube for cracks or holes d Replace quick disconnect fitting 4 Solenoid failu...

Page 80: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 5 13 Troubleshooting NOTES...

Page 81: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 5 14 Troubleshooting This page intentionally left blank...

Page 82: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 6 1 Drawings Millenium RO System CHAPTER SIX DRAWINGS...

Page 83: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 6 2 Drawings This page intentionally left blank...

Page 84: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 6 3 Drawings Millenium RO Component Layout...

Page 85: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 6 4 Drawings Millenium RO Disinfect View...

Page 86: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 6 5 Drawings Millenium RO Flow Schematic...

Page 87: ...ET PUMP MOTOR FAN 4 3 2 1 5 F3 7 8 9 10 6 1 POWER NOT APPLIED NOTES F3 1 8 AMP SLO BLOW 50Hz F3 1 4 AMP SLO BLOW 60Hz F1 8 AMP SLO BLOW F2 5 AMP SLO BLOW FUSE CALLOUT INLET W FUSE POWER CORD 50Hz UNIT...

Page 88: ...Drawings ALARM SET SWITCH CONDUCTIVITY PC BOARD DISPLAY PC BOARD PRODUCT WATER TDS REJECTION PC BOARD SENSOR SENSOR SENSOR SONALERT CONDUCTIVITY INLET WATER PRODUCT WATER THERMISTOR POWER SUPPLY REJEC...

Page 89: ...GE Osmonics 14693e 31 Mar 03 6 8 Drawings 1234687 50Hz 12315 60HZ 12315 1 50HZ ELECTRICAL CONTROL BOX ELECTRICAL CONTROL PANEL 12311 30010 1236062 60Hz 12310 30005 20203 20204 20205 12312 21080 Millen...

Page 90: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 6 9 Drawings 40569 1234678 Millenium RO Input Manifold 50 60Hz...

Page 91: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 6 10 Drawings 20178 40602 12131 14765 40569 40564 40218 Millenium RO Main Manifold...

Page 92: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 6 11 Drawings 40570 40218 40976 40976 40976 40976 40976 40218 14738 Millenium RO Waste Manifold...

Page 93: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 6 12 Drawings NOT TO SCALE DETAIL A OS1151581 Millenium RO ROMA Assembly...

Page 94: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 6 13 Drawings TO HOLE IN RO CABINET 41566 50112 41566 42014 14493 Millenium RO Flush Valve Assembly...

Page 95: ...WISE TO DECREASE PRESSURE EACH 90 ROTATION WILL CHANGE PRESSURE ABOUT 25 PSI WASHER PUMP ACORN NUT INCREASE PRESSURE ADJ SCREW DECREASE PROCEDURE TOOLS REQUIRED 5 REPLACE WASHER IF IT CAME OFF AND ACO...

Page 96: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 6 15 Drawings Push In Fitting Instructions...

Page 97: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 6 16 Drawings This page intentionally left blank...

Page 98: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 6 17 Drawings NOTES...

Page 99: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 6 18 Drawings This page intentionally left blank...

Page 100: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 7 1 Spare Parts List Millenium RO System CHAPTER SEVEN SPARE PARTS LIST...

Page 101: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 7 2 Spare Parts List This page intentionally left blank...

Page 102: ...RO only use w P n 1236173 1236173 Plug End 50Hz RO only use w P n 30382 42048 Gauge Product Pressure 42074 Gauge Pump Pressure 40179 Membrane Tape Wrapped for 500 GPD RO 2 required OS1228070 Membrane...

Page 103: ...Female 40568 Elbow speed fit 3 8 x 3 8 mnpt swivel 40564 Elbow speed fit 3 8 x 3 8 stem 41566 Elbow speed fit 1 4 x 3 8 stem 40742 Elbow speed fit 1 4 x 1 4 stem 40569 Elbow speed fit 3 8 x 1 4 mnpt...

Page 104: ...Spare Parts List 12128 Pre filter Housing Assembly 40694 O ring Pre filter 9 7 8 40695 Filter 1 Micron 9 7 8 40696 Filter 5 Micron 9 7 8 40698 Filter Carbon 9 7 8 14924 Kit Mil Prefilter Install Mount...

Page 105: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 7 6 Spare Parts List NOTES...