5 Maintenance
5.3 Inserting a new syringe
50
MicroCal iTC
200
Operating Instructions
28-9639-77
AA
3
Insert the syringe storage case into the loading hole.
Figure 5-1.
Syringe removal.
4
Slide the pipette to the cleaning position and insert it firmly into the hole.
5
Turn the inner metal syringe holder to the right several full rotations. This will
unscrew the metal nut that holds the syringe in the pipette.
6
Once the nut is loose, move the pipette to the loading hole.
7
Use the manual pipette controls to lower the pipette tip about 0.3 inches. Without
the bottom nut, the syringe will move down with the pipette tip.
8
Pull down on the syringe, gently, until it slides down into the storage case. The soft
grip tweezers may help with this.
9
Move the pipette out of the way, and screw the cap onto the storage case.
10 Remove it from the loading hole.
5.3
Inserting a new syringe
1
Insert the storage case into the loading hole and unscrew the top.
2
Position the pipette over it.