Installation 3
Site requirements 3.1
3 Installation
This section provides information about the installation of MicroCal iTC
200
.
Figure 3-1.
MicroCal iTC
200
with washing module and control computer..
Any equipment connected to the MicroCal iTC
200
must fulfill applicable standards and
local regulations.
3.1
Site requirements
The MicroCal iTC
200
with Controller requires about 1 meter of normal bench space (ca.
70 cm wide). This location should be away from strong drafts, room temperature
fluctuations, intense sunlight, vibrations and strong electrical or magnetic fields (as
may be produced by an NMR, microwave oven, large motors or refrigeration units). In
addition the mains power source (100 to 240 VAC) should be properly grounded and free
from voltage fluctuations, harmonic distortions, power dips and spikes. The AC power
line should be dedicated to the MicroCal iTC
200
system and should not share that power
with additional equipment.
NOTICE
It is recommended that the installation of the MicroCal iTC
200
instrument
be performed by GE Healthcare personnel.