Operation in the Vehicle Mounting Unit, Mounting the AC Power Adapter
227 490 02-C
Marquette Responder® 3000
73
16 Operation in the Vehicle Mounting Unit,
Mounting the AC Power Adapter
Operating the Defibrillator in the Vehicle
Mounting Unit
When you place the defibrillator in the mounting
unit, it will be powered from the ambulance power
system. Check that the power supply contacts on
the defibrillator and on the mounting unit (Figure
16-1) are clean. Always secure the defibrillator
with both locking devices (Figure 16-1, arrows at
top).
In the vehicle mounting unit, the defibrillator
display is upside down. Therefore you can flip the
display (see chapter 3, section "Display Flip").
Figure 16-1. Vehicle mounting unit
Mounting the AC Power Adapter
The AC power adapter is screwed to the rear of the
Marquette Responder
3000.
•
Remove the two legs (a, Figure 16-2) and the
screws (b) on the underside.
•
Attach the AC power adapter and secure it with
the new screws supplied.
•
Connect the device to the power line.
•
Check that the batteries are charging.
•
Remove the batteries and check the device
performance (chapter 3 "Putting the Defibril-
lator Into Operation").
Note
On line power, the defibrillator charges to 360
Joules in approx. 9 seconds.
a
b
a
Figure 16-2. Mounting the AC power adapter