Technical Specifications
82
Marquette Responder® 3000
227 490 02-C
After recording the ECG, the recorder will print an
intervention report (name, date of birth, user,
comments – entered by the user - defibrillation
mode, selected and delivered energy, number of
shocks and alarm limits).
Direct writing with rectangular coordinates using
thermorecording technology (printhead with
electronically controlled thermal elements records
on thermosensitive paper), roll paper, paper
transport by stepper motor:
•
number of recorder channels: 3
•
paper width: 90 mm
•
max. roll diameter 55 mm (roll with 40 m of
HELLIGE CONTRAST® paper)
•
printhead resolution: vertically 8 dots/mm,
horizontally 40 dots/mm at 25 mm/s
•
recorder speed 25 mm/s ±5 %
•
end-of-paper detection and message on display
To prevent damage to the printhead use only the
original HELLIGE CONTRAST paper.
Memory
•
storage of 180 minutes of ECG data with full
intervention report
•
storage of three 12SL ECG analysis reports
including the ECG
•
1 and 4-hour trends for ECG and etCO
2
•
storage of the most recent 250 actions (e.g.
device on/off, alarms, defibrillation energy)
including date and time
Pacemaker
•
Operating Modes: Demand, Fix
•
pacer rate: 30 to 200 P/min +-5%
•
pacer current: 0 to 200 mA +- 10% (for 500
Ω
)
•
pulse width: 20 ms
•
pulse shape: monophase square wave pulse
SpO
2
•
saturation: 0 to 100% in increments of 1 %
•
rate: 0 to 250 P/min in increments of 1 P/min
•
limit value: OFF, 60 to 100%
•
display of the plethysmogram
•
C-Lock ECG synchronization
•
Integration time: 4, 8, and 12 seconds
•
measurement error: 70 to 100% ±2 digits
50 to 69% ±3 digits
•
pulse readout 1.2% or ±1 P/min
etCO
2
semi-quantitative measuring method, using IR
technology
This method is based on the condition that the
inspired air is free of CO
2
.
•
measuring range: 0 to 76 mmHg
•
error (at ambient pressure of 1 atmosphere and
no CO
2
in the inspired air:
±
4 mmHg (
≤
40
mmHg);
±
10% (
>
40 mmHg)
•
response time: approx. 200 milliseconds
•
rise time: approx. 100 ms (10 to 90%)
•
warm-up time : approx. 10 seconds
•
dimensions: 58H
×
39W
×
17D mm (w/o.
airway adapter, cable and CO
2
sensor)
•
weight: 100 g (incl. CO
2
sensor)
•
safety standard IEC 60601-1, type BF
•
capnogram display