Semiautomatic Defibrillation
227 490 02-C
Marquette Responder® 3000
35
R
L
C
N
F
red
white
black
yellow
green
Figure 5-6.
ECG electrode placement
Check patient
ECG
I
15.07.1999 09:05:00
160 / 40
bpm
Energy
available
200 J
161
ECG
SpO2
etCO2
QRSPulse
Tone
OFF
Next
Menu
II
III
Figure 5-7.
ECG display (acquisition via
separate ECG electrodes)
Note
The main menu will automatically reappear, when
no button is activated for a period of 30 seconds.
ECG Acquisition via Separate ECG Electrodes
Use only silver/silver-chloride electrodes to
acquire the ECG signal. These electrodes prevent
polarization voltages which may be caused by the
defibrillation shock, resulting in an ECG trace on
the monitor screen or recording that simulates
cardiac arrest. The ECG can be picked up with 5
or with 10 electrodes (for ECG measurement,
however, 10 ECG electrodes are required (chapter
8)).
•
Apply the electrodes as shown in Figure 5-6.
•
Ensure that only those leadwires are connected
to the patient cable that are actually required.
•
Via the patient cable, connect the electrodes to
the defibrillator.
•
Turn on the device (energy selector to
).
On the monitor screen you will now see the 3 ECG
leads selected in the setup menu (factory defaults:
leads I, II, III, Figure 5-7). The analysis algorithm
always uses ECG lead II. If lead II is not available,
the first suitable ECG lead shown will be selected.
Follow these steps to change the displayed leads or
the signal size:
−
Press F1
ECG
.
−
Press F1
I...III, aV...V, Paddle
again.
A menu for selection of the ECG lead ap-
pears.
−
Using F1, select the lead to be displayed in
channel 1, with F2 select the lead for chan-
nel 2, and with F3 select the lead for chan-
nel 3 (with each key press the device ad-
vances to the next lead).
−
Press F5
Previous Menu
and change the signal
size with F2
1 cm/mV
.
−
Press F5
Next Menu
for about 2 seconds to
return to the main menu.
•
Defibrillate the patient as described on page 32.