5-32
MAC™ 3500 Resting ECG Analysis System
Revision L
2021337-036
Maintenance:
COMM Board Replacement
10. Select
Return
and press
Enter
.
The
Transmission
window closes.
11. On the
Basic System
menu, select
Return
and press
Enter
.
The
Basic System
menu closes.
12. On the
System Setup
menu, select
Save Setup
>
To System
.
This saves the settings to the communication board and displays the following
message:
This should take approximately 16 to 19 seconds. If it takes less than 16
seconds, the settings were not saved successfully. Repeat a second time.
13. Wait at least 10 minutes and verify the MAC device can communicate with the
MUSE system.
Windows File Server 2000/2003 has an ARP cache that regenerates the MAC
Address/Physical Address table every 10 minutes. When you replace the
communications board, you change the MAC Address/Physical Address. By
waiting at least 10 minutes, you allow the MUSE operating system to
regenerate the MAC Address/Physical Address cache so the MAC device is
able to communicate again with the MUSE system.
NOTE
If the MAC device cannot communicate with the MUSE system after 10
minutes, cycle the power on the MAC device and wait another 10 minutes
to reverify the device can communicate with the MUSE system. Cycling
the power ensures the network settings are reread from the communications
board.
14. Perform all necessary functional checkouts.
See
“Functional Checkout Procedures”
Please wait.
Summary of Contents for MAC 3500
Page 9: ...Revision L MAC 3500 Resting ECG Analysis System 1 1 2021337 036 1 Introduction ...
Page 10: ...1 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Page 19: ...Revision L MAC 3500 Resting ECG Analysis System 2 1 2021337 036 2 Equipment Overview ...
Page 20: ...2 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Page 25: ...Revision L MAC 3500 Resting ECG Analysis System 3 1 2021337 036 3 Installation ...
Page 26: ...3 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Page 41: ...Revision L MAC 3500 Resting ECG Analysis System 4 1 2021337 036 4 Troubleshooting ...
Page 42: ...4 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Page 83: ...Revision L MAC 3500 Resting ECG Analysis System 5 1 2021337 036 5 Maintenance ...
Page 84: ...5 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Page 123: ...Revision L MAC 3500 Resting ECG Analysis System 6 1 2021337 036 6 Parts Lists ...
Page 124: ...6 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Page 153: ...Revision L MAC 3500 Resting ECG Analysis System A 1 2021337 036 A Technical Specifications ...
Page 154: ...A 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Page 162: ...B 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Page 168: ...C 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Page 175: ...Revision L MAC 3500 Resting ECG Analysis System Index 1 2021337 036 Index ...
Page 176: ...Index 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Page 178: ...Index Index 4 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 ...
Page 179: ......