4-16
MAC™ 3500 Resting ECG Analysis System
Revision L
2021337-036
Troubleshooting:
Diagnostic Tests
Battery Test
Battery tests check the current battery status, battery discharge rate and battery
charge rate. Test results are stored in memory and can be printed out. The graphic
displayed shows the
Battery Test
menu. Each test is covered in detail below.
Battery Status
. This test displays and constantly updates current information on
the battery voltage, battery current, percent of charge remaining and battery
temperature.
Battery Voltage
. With a reading of 80% or more for the percent of charge
remaining, the battery voltage should be between 15 and 24 volts. If battery
voltage is below 15 volts, the battery may need to be replaced.
Battery Current
. Disconnect AC power. If the battery current is less than -
0.7 amps, the main CPU may need to be replaced. For example if current is
-0.8 amps, consider replacing CPU. See
Battery Temperature
. A temperature reading over 45° C indicates a failure.
If battery temperature is more than 10° C over the ambient temperature
consider replacing battery.
Ambient Temperature
. Indicates the temperature inside the unit. The
temperature displayed is accurate to 5°C in the range of 0°C to
50°C. Ambient temperatures not within this range cannot be displayed.
19A
20A
Summary of Contents for MAC 3500
Page 9: ...Revision L MAC 3500 Resting ECG Analysis System 1 1 2021337 036 1 Introduction ...
Page 10: ...1 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Page 19: ...Revision L MAC 3500 Resting ECG Analysis System 2 1 2021337 036 2 Equipment Overview ...
Page 20: ...2 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Page 25: ...Revision L MAC 3500 Resting ECG Analysis System 3 1 2021337 036 3 Installation ...
Page 26: ...3 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Page 41: ...Revision L MAC 3500 Resting ECG Analysis System 4 1 2021337 036 4 Troubleshooting ...
Page 42: ...4 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Page 83: ...Revision L MAC 3500 Resting ECG Analysis System 5 1 2021337 036 5 Maintenance ...
Page 84: ...5 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Page 123: ...Revision L MAC 3500 Resting ECG Analysis System 6 1 2021337 036 6 Parts Lists ...
Page 124: ...6 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Page 153: ...Revision L MAC 3500 Resting ECG Analysis System A 1 2021337 036 A Technical Specifications ...
Page 154: ...A 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Page 162: ...B 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Page 168: ...C 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Page 175: ...Revision L MAC 3500 Resting ECG Analysis System Index 1 2021337 036 Index ...
Page 176: ...Index 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Page 178: ...Index Index 4 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 ...
Page 179: ......