5-38
MAC™ 3500 Resting ECG Analysis System
Revision L
2021337-036
Maintenance:
Functional Checkout Procedures
Electrical Safety Checks
20. Current leakage and ground continuity test results meet requirements?
Perform electrical safety checks when indicated. All indicated electrical safety
checks require a pass/fail indication for steps performed. Record the
measurement values in your debrief.
Table 6. Electrical Safety Checks
Step
1
Condition
2
UUT - ON
Result
Leakage
Current Limits
Earth Leakage Current
1.
Forward Polarity
NC
______
Pass/Fail
300
2.
Neutral open, Forward Polarity
SFC
______
Pass/Fail
1,000
3.
Neutral open, Reverse Polarity
SFC
______
Pass/Fail
1,000
4.
Reverse Polarity
NC
______
Pass/Fail
300
Enclosure Leakage Current
1.
Forward Polarity
NC
______
Pass/Fail
100
2.
Neutral open, Forward Polarity
SFC
______
Pass/Fail
300
3.
Ground open, Forward Polarity
SFC
______
Pass/Fail
300
4.
Ground open, Reverse Polarity
SFC
______
Pass/Fail
300
5.
Neutral open, Reverse Polarity
SFC
______
Pass/Fail
300
6.
Reverse Polarity
NC
______
Pass/Fail
100
Patient Leakage Current to Ground
1.
Forward Polarity
NC
______
Pass/Fail
10
2.
Neutral open, Forward Polarity
SFC
______
Pass/Fail
50
3.
Ground open, Forward Polarity
SFC
______
Pass/Fail
50
4.
Ground open, Reverse Polarity
SFC
______
Pass/Fail
50
5.
Neutral open, Reverse Polarity
SFC
______
Pass/Fail
50
6.
Reverse Polarity
NC
______
Pass/Fail
10
Patient Leakage Current Mains on Applied Part
3
1.
Forward Polarity
Neutral / Ground Closed
SFC
______
Pass/Fail
5000
2.
Reverse Polarity
Neutral / Ground Closed
SFC
______
Pass/Fail
5000
Summary of Contents for MAC 3500
Page 9: ...Revision L MAC 3500 Resting ECG Analysis System 1 1 2021337 036 1 Introduction ...
Page 10: ...1 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Page 19: ...Revision L MAC 3500 Resting ECG Analysis System 2 1 2021337 036 2 Equipment Overview ...
Page 20: ...2 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Page 25: ...Revision L MAC 3500 Resting ECG Analysis System 3 1 2021337 036 3 Installation ...
Page 26: ...3 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Page 41: ...Revision L MAC 3500 Resting ECG Analysis System 4 1 2021337 036 4 Troubleshooting ...
Page 42: ...4 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Page 83: ...Revision L MAC 3500 Resting ECG Analysis System 5 1 2021337 036 5 Maintenance ...
Page 84: ...5 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Page 123: ...Revision L MAC 3500 Resting ECG Analysis System 6 1 2021337 036 6 Parts Lists ...
Page 124: ...6 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Page 153: ...Revision L MAC 3500 Resting ECG Analysis System A 1 2021337 036 A Technical Specifications ...
Page 154: ...A 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Page 162: ...B 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Page 168: ...C 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Page 175: ...Revision L MAC 3500 Resting ECG Analysis System Index 1 2021337 036 Index ...
Page 176: ...Index 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Page 178: ...Index Index 4 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 ...
Page 179: ......