Maintenance
The following window opens:
8.
If the
WARNING
window opens, press
F6
(
OK
), and continue with the software
update.
9.
Perform the applicable checkout procedures.
Refer to
“Functional Checkout” on page 99
.
Conditioning The Battery Pack
To maintain the storage capacity of the battery pack installed in your device, GE
Healthcare recommends that you condition the device’s battery pack once every
six months to reset the electronic fuel gauge inside the battery. A condition cycle
consists of an uninterrupted “charge-discharge-charge” cycle.
You can condition the battery pack while it is installed in a device that you are not
currently using to record tests on a patient.
1.
Disconnect the AC power from the device.
2.
Display the
BATTERY STATUS
window:
a.
On the
Main Menu
, press
F5
(
System Configuration
).
b.
Press
F6
(
More
) >
F6
(
More
) >
F5
(
Service Setup
).
c.
When prompted, enter the
Service password
and press
F6
(
OK
).
If you do not know the service password, contact GE Healthcare Technical
Support.
d.
Select
System Diagnostics
.
e.
Select
Battery Test
.
The
BATTERY STATUS
window opens.
3.
Allow the battery to discharge until
Battery Charge Remaining [%]
is less than
90%.
4.
Turn off the device and reconnect it to the AC power.
106
MAC™ 1600
2028451-183E
Summary of Contents for MAC 1600
Page 18: ...Introduction 18 MAC 1600 2028451 183E...
Page 26: ...Equipment Overview Block Diagram 26 MAC 1600 2028451 183E...
Page 28: ...Equipment Overview 28 MAC 1600 2028451 183E...
Page 108: ...Maintenance 108 MAC 1600 2028451 183E...
Page 111: ...Parts Lists 2028451 183E MAC 1600 111...
Page 112: ...Parts Lists 112 MAC 1600 2028451 183E...
Page 113: ...Parts Lists 2028451 183E MAC 1600 113...
Page 128: ...Parts Lists 128 MAC 1600 2028451 183E...
Page 140: ...Electromagnetic Compatibility 140 MAC 1600 2028451 183E...
Page 145: ......