Basic Service Manual
Section 8-8 - Replacing Main Console
8 - 159
8-8-26 Replacement of the Unity Rear Top Fan Assy
This table includes information specific to these instructions. For more information
for servicing the LOGIQ Totus" on page 8-5
8-2-7 "PPE Required During Service" on page 8-6
8-8-26-1
Preparations
1.) Remove the T2 Side Cover R Assy as described in
8-8-8 "Replacement of the T2 Side Cover R
2.) Remove the Rear Cover Mid Assy-T as described in
8-8-10 "Replacement of the Rear Cover Mid
8-8-26-2
Removal Procedure
Table 8-88 Manpower, Time, Tools, and PPE
Manpower
Time
Tools
PPE
One person
Approximately 10 minutes
Phillips Screwdriver
Cut resistant gloves
Knee Pad
Step
Corresponding graphic
1. Disconnect the Fan Cable from the console.
2.
Unscrew 3 screws to remove the Unity Rear
Top Fan Assy from the console