Care and Maintenance
4-44
Ultrasound System
–
Common Service Information
Direction 5444964-100 English
Rev. 5
When there's too much leakage
current …
AC/DC Fails
Where applicable, check the AC/DC adapter and its cable.
Replace a new one if any portion is defective.
Chassis Fails
Check the ground on the power cord and plug for continuity.
Ensure the ground is not broken, frayed, or intermittent. Replace
any defective part.
Where applicable, tighten all grounds. Ensure star washers are
under all ground studs.
Inspect wiring for bad crimps, poor connections, or damage.
Test the wall outlet; verify it is grounded and is free of other
wiring abnormalities. Notify the user or owner to correct any
deviations. As a work around, check the other outlets to see if
they could be used instead.
NOTE:
No outlet tester can detect the condition where the white neutral
wire and the green grounding wire are reversed. If later tests
indicate high leakage currents, this should be suspected as a
possible cause and the outlet wiring should be visually
inspected. If reversed wiring is observed, it is the customers
responsibility to have the condition corrected by qualified
personnel.
Summary of Contents for LOGIC C2 Series
Page 2: ...2013 2018 General Electric Company Manufacturer GE Medical Systems www gehealthcare com...
Page 6: ...i 4 Ultrasound System Common Service Information Direction 5444964 100 English Rev 5...
Page 7: ...Ultrasound System Common Service Information i 5 Direction 5444964 100 English Rev 5...
Page 8: ...i 6 Ultrasound System Common Service Information Direction 5444964 100 English Rev 5...
Page 9: ...Ultrasound System Common Service Information i 7 Direction 5444964 100 English Rev 5...
Page 10: ...i 8 Ultrasound System Common Service Information Direction 5444964 100 English Rev 5...
Page 11: ...Ultrasound System Common Service Information i 9 Direction 5444964 100 English Rev 5...
Page 12: ...i 10 Ultrasound System Common Service Information Direction 5444964 100 English Rev 5...
Page 13: ...Ultrasound System Common Service Information i 11 Direction 5444964 100 English Rev 5...
Page 14: ...i 12 Ultrasound System Common Service Information Direction 5444964 100 English Rev 5...
Page 15: ...Ultrasound System Common Service Information i 13 Direction 5444964 100 English Rev 5...
Page 20: ...i 18 Ultrasound System Common Service Information Direction 5444964 100 English Rev 5...
Page 24: ...i 22 Ultrasound System Common Service Information Direction 5444964 100 English Rev 5...
Page 127: ...2013 2018 General Electric Company Manufacturer GE Medical Systems www gehealthcare com...