6 Reference and Specification
K0470 Revision A
6 - 30
[EN] English
Volts-free Contact Option
The Volts-free Contact option provides a selectable relay contact toggle depending on conditions
set in the PACE instrument.
Pin number
Function
Pin number
Function
1
Relay 1 normally CLOSED
9
Relay 3 common
2
Relay 1 normally OPEN
10
0V return
3
Relay 1 common
11
+24V DC OUT @ 100mA
4
Relay 2 normally CLOSED
12
SW IN 1
5
Relay 2 normally OPEN
13
SW IN 2
6
Relay 2 common
14
not used
7
Relay 3 normally CLOSED
15
not used
8
Relay 3 normally OPEN
Each selection has
three Volts-free
Contacts.
Selection of trigger
conditions.
Selection of trigger
conditions.
30V maximum with respect to
chassis.
Rated output = 24V
30V
max
To maintain PACE product safety, external
circuits connected to the instrument must meet
Safety Extra-Low Voltage (SELV) requirements.
0V
+V
event
trigger
24 Vdc maximum
30 Vdc wrt chassis
1 A resistive maximum
Typical Volts-free
schematic
Relay contacts rated at 30 Vdc, 1 A resistive,
200 mA inductive.
22V to 26V
ANALOGUE 1
ANALOGUE 2
15
CONSULT HANDBOOK
volts-free connections
Summary of Contents for K0470