6 Reference and Specification
K0470 Revision A
6 - 28
[EN] English
6.10 Options
Option enable process
To enable soft options on a PACE instrument, use the following:
1.
Touch the top Measure area of the screen.
2.
Select Global Setup.
3.
Select Calibration.
4.
Enter a Calibration PIN 1234.
5.
Enter new option key xxxxxxxxxx (10 digits).
6.
After entry of this key PACE confirms the options have been enabled.
Note:
Hardware options automatically enable post installati
on
Selects Analogue
Output Range
On/Off
Update rate of Analogue Output
option from the control module.
Analogue Output Bandwidth = 0.5 x Update Rate (Hz)
30V maximum with respect to
chassis.
Rated output = 24V
30V
max
To maintain PACE product safety, external
circuits connected to the instrument must meet
Safety Extra-Low Voltage (SELV) requirements.
15-way female D connector
Front view of Analogue
Option PCB connector
1
2
3
4
5
6
7
8
15
14
13
12
11
10
9
30V
max
22V to 26V
ANALOGUE 1
ANALOGUE 2
15
CONSULT HANDBOOK
analogue connections