Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare
Direction 5314528-2-1EN, Revision 1
Relative Humidity and Temperature: Refer to
. To obtain relative humidity and
temperature requirements for components not specified in
, refer to the appropriate
component Pre–Installation Manual listed in
Basic Innova LC System
Compatibility.
Altitude and Atmospheric Pressure: Refer to
. To obtain altitude and
atmospheric pressure requirements for components not specified in
, refer to the
appropriate component Pre–Installation Manual listed in
Basic Innova
LC System Compatibility.
Table 3-2:
INSTALLATION ROOM
OF PRODUCT OR
COMPONENT
ALTITUDE (meters)
ATMOSPHERIC PRESSURE (kPa)
IN-USE
STORAGE /
TRANSPORT
IN-USE
STORAGE / TRANSPORT
MIN
MAX
MIN
MAX
MIN
MAX
MIN
MAX
Examination room
0
3000
-30
3000
70
106
70
106
Technical room (See
Notice below
0
3000
-30
3000
70
106
50
106
Control room
0
3000
0
3000
N/A
N/A
N/A
N/A
The chiller is able to dissipate maximum continuously power at 3000 meters in an
ambient temperature up to 20 C, for this altitude the technical room temperature
shall not exceed 20 C.
In some cases condensation occurs and water drops from outlets and pipes of the
air conditioner in the technical room.
Therefore, it is critical to install the cabinets where there is no risk of flood from the
air conditioner.
Ensure the air outlet positions in the exam room is not in area closed to monitor
suspension carriage/rail.
2.2 Equipment Heat Output (Dissipation)
2.2.1 Equipment Heat Output tables
Refer to
. To obtain heat output information for components not specified in
3-3
, refer to the appropriate component Pre–Installation Manual listed in
3,
Basic Innova LC System Compatibility.
Chapter 3 Physical Requirements of Room
47