Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare
Direction 5314528-2-1EN, Revision 1
Table 3-3:
HEAT OUTPUT
Stand by
Moderate Use (4)
Typical Use (4)
Maximum Use (4)
Room
Core System
kW
BTU/hr
kW
BTU/hr
kW
BTU/hr
kW
BTU/hr
Exam
Room
Innova LC positioner and
table
0.41
1394
0.55
1858
0.89
3020
1.62
5517
DL user area with 1 TFT
monitor
0.16
546
0.16
546
0.16
546
0.16
546
Ctrl Room
2 B&W flat monitors
0.17
573
0.17
573
0.17
573
0.17
573
C1 Cabinet
0.41
1398
0.69
2366
0.99
3389
1.29
4412
C2 Cabinet
0.29
989
0.33
1125
0.54
1828
0.87
2966
Coolix X-Ray tube chiller
(1) (2)
2.53
8619
4.49
15309
5.49
18725
6.93
23625
Detector conditioner
0.21
709
0.21
709
0.21
709
0.21
709
Tech
Room
Main disconnect panel -
PDB
0.4
1534
0.45
1534
0.45
1534
0.45
1534
Total for core system
4.6
15762
7.0
24020
8.9
30324
11.7
39881
Room
Options (3 & 5)
Stand by
Moderate Use (4)
Typical Use (4)
Maximum Use (4)
3 in room B&W TFT
monitors
0.25
859
In room AW TFT monitor
0.12
409
Exam
Room
Typical injector
0.09
320
AW work station
0.35
1201
2 AW TFT monitors
0.24
818
Ctrl Room
Printer
0.31
1054
Same values as Stand
by
Same values as Stand
by
Same values as Stand
by
Typical configuration without fluoro UPS
5.9
20424
8.4
28682
10.3
34985
13.1
44543
NOTE:
(1) Air flow requirements 1200 m3/h (706 CFM)
NOTE:
(2) For more details, consult appropriate pre–installation manual
NOTE:
(3) For UPS 20 kVA option refer to
NOTE:
(4) Moderate use: 8 cases / 10 hours, typical use: 11 cases / 10 hours, maximum use:
during case.
NOTE:
(5) For IVUS enviromental limatations and heat output, refer to
48
2 Environmental Requirements/Limitations