Caution
5
4
Caution
Sample Size Example
0.75µL 1.0
1.5µL 2.0µL
µL
3.0µL
- Avoid contact with dripping or splashing liquids.
- The minimum blood sample size to test using the GE200 Blood Glucose Monitoring System is 0.75µL: ( )
Blood sample size above 3.0
µL
might
contaminate the
test strip port and the meter
.
Blood sample size below 0.75µL
m
ay cause an inaccurate result or may prevent a meter reading
.
An Er4 reading will
be displayed if the sample size is too small.
In this case, repeat the test with a new test strip.
Important Safety Notes:
- The GE200 Blood Glucose Meter and Lancing Device are for single patient use. Do not share them with
anyone including other family members.
- All parts of the kit are considered biohazardous and can potentially transmit infectious diseases, even after
following the cleaning and disinfecting procedures. Please refer to the section "Cleaning and Disinfecting
Procedures" on page 47.
- Users should wash their hands thoroughly with soap and water before and after handling the meter,
lancing device, or test strips.