About Quality Control Testing
Understanding Test Results and Messages
Control Solution Range:
Example of Control Solution Range printed on
your test strip vial label.
Control Solution Range mmol/L
Normal (Level 2):
High (Level 4):
4.9-6.7
13.8-18.8
The
Meter displays results between
If the test result is below 1.1 mmol/L, “ ” will appear on the screen. Please repeat your test
with by a new test strip.
If you still get a “ ” result, contact your healthcare professional.
If the test result is above 33.3 mmol/L, “ ” will appear on the screen. Please repeat your
test again with a new test strip.
If you still get a “ ” result, contact your healthcare professional.
GE200 Blood Glucose
1.1 and 33.3 mmol/L.
What is a Quality Control Test ?
To ensure proper monitoring function, it is necessary to regularly perform a quality control test.
Use one of the GE200 Control Solutions when testing your GE200 Blood Glucose Monitoring System in the Control
Solution Mode. If the test result is within the Control Solution Range printed on the strip vial label, the GE200 Blood
Glucose Monitoring System passes the Quality Control Test and your GE200 Blood Glucose Monitoring System is
working properly.
NOTE
- If you want to purchase Level 2 (Normal) control solution, please contact the
toll free at
Auto Control Medical
Distribution and Customer Service Center at 1-800 964-8434
1-800 964 8434.