[EN] English -
11
K393 Issue 3
Calibration - Procedures (mV or Volts output)
1. Connect the instrument to the calibration equipment
2. Let the equipment get to a stable temperature
(minimum: 5 minutes since the last power on).
3. Use the calibration menu (Table 9) to do a two-point
calibration (Zero and +FS). The display shows the
applicable instructions to complete the calibration.
4. To make sure that the calibration is correct, select the
applicable
Volts
output task and range (Table 2).
5. Then set the output values that are applicable to the
calibration:
• mV: 0, 30, 60, 90, 120
OR
• Volts (V): 0, 3, 6, 9, 12
6. Make sure that the error is in the specified limits
or
)
.
Table 13:
mV output error limits
Table 14:
Volts (V) output error limits
Calibration - Procedures (mA output)
1. Connect the instrument to the calibration equipment
2. Let the equipment get to a stable temperature
(minimum: 5 minutes since the last power on).
3. Use the calibration menu (Table 9) to do a two-point
calibration (Zero and +FS). The display shows the
applicable instructions to complete the calibration.
4. To make sure that the calibration is correct, select the
applicable mA output task (Table 2) and set these
output values:
• mA: 0.1, 4, 12, 20, 24
5. Make sure that the error is in the specified limits
.
Table 15:
mA output error limits
Calibration - Procedures (IDOS UMM)
Refer to the user manual for the IDOS UMM.
When the calibration is complete, the instrument
automatically sets a new calibration date in the UMM.
Output
mV
Calibrator
error
(mV)
Permitted
DPI 832 error
(mV)
0
0.00005
0.01
30
0.000425
0.02
60
0.0008
0.03
90
0.001175
0.03
120
0.00098
0.04
Output
V
Calibrator
error
(V)
Permitted
DPI 832 error
(V)
0
0.00000005
0.001
3
0.0000175
0.002
6
0.00003
0.002
9
0.00005
0.002
12
0.000134
0.002
Output
mA
Calibrator
error
(mA)
Permitted
DPI 832 error
(mA)
0.1
0.000006
0.001
4
0.00020
0.001
12
0.0014
0.001
20
0.00185
0.002
24
0.0023
0.002
Summary of Contents for Druck DPI 832
Page 2: ......
Page 6: ...Druck DPI 832 Electrical loop calibrator User manual K393 EN DE ES FR IT PT CN JP RU ...
Page 7: ......
Page 20: ...Druck DPI 832 Kalibrator für Elektrische Signale Handbuch K393 EN DE ES FR IT PT CN JP RU ...
Page 21: ......
Page 34: ...Druck DPI 832 Calibrador de lazo eléctrico Manual del usuario K393 EN DE ES FR IT PT CN JP RU ...
Page 35: ......
Page 49: ......
Page 62: ...Druck DPI 832 Calibratore di loop elettrico Manuale utente K393 EN DE ES FR IT PT CN JP RU ...
Page 63: ......
Page 76: ...Druck DPI 832 Calibrador de loop elétrico Manual do usuário K393 EN DE ES FR IT PT CN JP RU ...
Page 77: ......
Page 90: ...Druck DPI 832 电气回路校验仪 用户手册 K393 EN DE ES FR IT PT CN JP RU ...
Page 91: ......
Page 104: ...Druck DPI 832 電気ループ校正器 ユーザーマニ ュ アル K393 EN DE ES FR IT PT CN JP RU ...
Page 105: ......
Page 117: ......
Page 131: ......