K393 Issue 3
Selection sequence:
➤
Select Task
➤
Set Up
➤
Calibration
➤
.
Table 9:
Calibration options
When you select a channel, the display shows the
applicable instructions to complete the calibration.
When the calibration is complete, select
Calibration Due
and set the new calibration date for the instrument.
Calibration - Procedures (mV or Volts input)
1. Connect the instrument to the calibration equipment
2. Let the equipment get to a stable temperature
(minimum: 5 minutes since the last power on).
3. Use the calibration menu (Table 9) to do a three-point
calibration (-FS, Zero and +FS). The display shows the
applicable instructions to complete the calibration.
4. To make sure that the calibration is correct, select the
applicable
Volts
input task and range (Table 2).
5. Then apply the input values that are applicable to the
calibration:
• mV: -120, -60, -30, 0 (short circuit)
Then mV: 0, 30, 60, 120
OR
• Volts (V): -30, -15, -5, 0 (short circuit)
Then volts (V): 0, 5, 15, 30
6. Make sure that the error is in the specified limits
Table 10:
mV input error limits
Table 11:
Volts (V) input error limits
Calibration - Procedures (mA input)
1. Connect the instrument to the calibration equipment
2. Let the equipment get to a stable temperature
(minimum: 5 minutes since the last power on).
3. Use the calibration menu (Table 9) to do a three-point
calibration (-FS, Zero and +FS). The display shows the
applicable instructions to complete the calibration.
4. To make sure that the calibration is correct, select the
applicable mA input task (Table 2) and apply these
values:
• mA: -55, -40, -24, -18, -12, -6, 0 (open circuit)
Then mA: 0, 6, 12, 18, 24, 40, 55.
5. Make sure that the error is in the specified limits
)
.
Table 12:
mA input error limits
1
Display: Enter
Calibration PIN
2
3
Menu:
Select Channel
4
5
▲
▼
+
■
■
▲
▼
Factory PIN = 4321
Options
Description
...
To calibrate the specified input: mV, Volts, mA
...
To calibrate the specified output: mA, mV, Volts
IDOS ...
UMM only. To calibrate the specified IDOS UMM.
Refer to the user manual for the IDOS UMM.
Calibration Due:
To set the date of the next
calibration for the instrument.
After the specified calibration date, there is a
warning message. There is a selection box to stop
the warning.
To change the calibration PIN (Personal
Identification Number).
Applied
mV
Calibrator
error
(mV)
Permitted
DPI 832 error
(mV)
±120
0.0013
0.03
±60
0.0008
0.02
±30
0.0006
0.02
0 (Short circuit)
-
0.01
Applied
V
Calibrator
error
(V)
Permitted
DPI 832 error
(V)
±30
0.00058
0.004
±15
0.00011
0.002
±5
0.00006
0.001
0 (Short circuit)
-
0.001
Applied
mA
Calibrator
error
(mA)
Permitted
DPI 832 error
(mA)
±55
0.0022
0.005
±40
0.0018
0.004
±24
0.0014
0.003
±18
0.0004
0.003
±12
0.0003
0.002
±6
0.0002
0.002
0 (open circuit)
-
0.001
Summary of Contents for Druck DPI 832
Page 2: ......
Page 6: ...Druck DPI 832 Electrical loop calibrator User manual K393 EN DE ES FR IT PT CN JP RU ...
Page 7: ......
Page 20: ...Druck DPI 832 Kalibrator für Elektrische Signale Handbuch K393 EN DE ES FR IT PT CN JP RU ...
Page 21: ......
Page 34: ...Druck DPI 832 Calibrador de lazo eléctrico Manual del usuario K393 EN DE ES FR IT PT CN JP RU ...
Page 35: ......
Page 49: ......
Page 62: ...Druck DPI 832 Calibratore di loop elettrico Manuale utente K393 EN DE ES FR IT PT CN JP RU ...
Page 63: ......
Page 76: ...Druck DPI 832 Calibrador de loop elétrico Manual do usuário K393 EN DE ES FR IT PT CN JP RU ...
Page 77: ......
Page 90: ...Druck DPI 832 电气回路校验仪 用户手册 K393 EN DE ES FR IT PT CN JP RU ...
Page 91: ......
Page 104: ...Druck DPI 832 電気ループ校正器 ユーザーマニ ュ アル K393 EN DE ES FR IT PT CN JP RU ...
Page 105: ......
Page 117: ......
Page 131: ......