8
- [EN] English
K393 Issue 3
Operation - Switch test
To do tests on a switch:
1. Connect the instrument (Figure 5) and, if necessary,
adjust the
Set Up
2. Select the applicable switch test from
Select Task
(Table 2/3) and, if necessary, adjust the
Settings
(Table 5). The display shows the switch condition (open
or closed) in the top right-hand corner.
3. Supply the output values to the system (Table 7).
• Example - “Nudge” output.
a. Use
Edit
(
■
■
) to set a value less than the switch
value.
b. Use the
▲
▼
buttons to change the value in small
increments.
• Example - “Ramp” output.
a. Set “High” and “Low” values that are applicable to
the switch value (Table 6). Then, to get an accurate
switch value, set a long “Travel” period.
b. Use
Start/Stop
(
■
■
) to start and stop the “Ramp”
cycle.
4. If necessary, supply the output values in the opposite
direction until the switch changes condition again.
The display shows the applicable values to open and
close the switch.
5. To do the test again, press
ESC
to reset the values.
Figure 5:
Example configuration - Switch test
Operation - UPM Pressure measurements
Read all the instructions supplied with the UPM and then
use the specified procedures to connect it (Figure 6/7).
Figure 6:
Example configuration - Pressure measurement
with a UPM
When the connections are complete, make the necessary
IDOS selections (Table 2/3).
If you re-attach a UPM, the instrument uses the same
measurement units that you used before. The instrument
keeps a record for the last 10 modules.
UPM - Measure the pressure
To measure the pressure (Figure 6):
1. Select the applicable pressure task from
Select Task
(Table 2/3) and, if necessary, adjust the
Set Up
and the
Settings
2. If necessary, do a zero correction (Table 4).
To measure pressure with another operation (Figure 7), use
the same procedure.
Edit
Settings
Settings
a) Pressure
b) Leak test
Start
Settings
Summary of Contents for Druck DPI 832
Page 2: ......
Page 6: ...Druck DPI 832 Electrical loop calibrator User manual K393 EN DE ES FR IT PT CN JP RU ...
Page 7: ......
Page 20: ...Druck DPI 832 Kalibrator für Elektrische Signale Handbuch K393 EN DE ES FR IT PT CN JP RU ...
Page 21: ......
Page 34: ...Druck DPI 832 Calibrador de lazo eléctrico Manual del usuario K393 EN DE ES FR IT PT CN JP RU ...
Page 35: ......
Page 49: ......
Page 62: ...Druck DPI 832 Calibratore di loop elettrico Manuale utente K393 EN DE ES FR IT PT CN JP RU ...
Page 63: ......
Page 76: ...Druck DPI 832 Calibrador de loop elétrico Manual do usuário K393 EN DE ES FR IT PT CN JP RU ...
Page 77: ......
Page 90: ...Druck DPI 832 电气回路校验仪 用户手册 K393 EN DE ES FR IT PT CN JP RU ...
Page 91: ......
Page 104: ...Druck DPI 832 電気ループ校正器 ユーザーマニ ュ アル K393 EN DE ES FR IT PT CN JP RU ...
Page 105: ......
Page 117: ......
Page 131: ......