   Safety 

K0454 Issue 1 - [EN]  English

3

Safety

Before you use the instrument, make sure that you read and 
understand all the related data. This includes: the applicable 
local safety procedures, the user manual (K0449), and the 
instructions for the accessories/options/equipment you are 
using it with.

General 
warnings

WARNING

It is dangerous to ignore the specified limits for the 
instrument or to use the instrument when it is not in its 
normal condition. Use the applicable protection and 
obey all safety precautions.

Do not use the instrument in locations with explosive 
gas, vapour or dust. There is a risk of an explosion.

Electrical 
warnings

To prevent electrical shocks or damage to the 
instrument, do not connect more than 30V between the 
terminals, or between the terminals and the ground 
(earth).

To prevent electrical shocks, use only the GE specified 
AC probe (Part: IO620-AC) to measure AC voltages 
(maximum: 300 Vac). 

This instrument uses a Lithium-Polymer (Li-Polymer) 
battery pack. To prevent an explosion or fire, do not 
short circuit, do not disassemble, keep it safe from 
damage. For operating conditions, see Table 1.

To prevent an explosion or fire, use only the GE 
specified battery, power supply and battery charger. 

To prevent battery leakage or heat generation, only use 
the battery charger and power supply in the 
temperature range 0 to 40°C (32 to 104°F). For 
operating conditions, see Table 1.

Pressure 
warnings

Some liquid and gas mixtures are dangerous. This 
includes mixtures that occur because of 
contamination. Make sure that the equipment is safe to 
use with the necessary media.

To prevent a dangerous release of pressure, isolate and 
bleed the system before you disconnect a pressure 
connection.

Continued

Summary of Contents for Druck DPI 620

Page 1: ...S Figure C4 k 300 Vac AC probe Figure C5 Part IO620 AC A1 2 DPI 620 Channel 2 CH2 l 30 V M m 2000 mV M n 55 mA M Figure D1 o 0 to 24 mA S p 24 V loop P q Switch M A1 3 DPI 620 MC 620 PM 620 Pressure...

Page 2: ...12 f c e 11 b a d b 2 a 1 a 2 2 1 1 3 3 5 4 a b 2 4 3 1 C1 DPI 620 CH1 C2 DPI 620 CH1 C3 DPI 620 CH1 C4 DPI 620 CH1 C5 DPI 620 CH1 E1 DPI 620 MC 620 PM 620 C 4 D1 DPI 620 CH2 A1 1 a b c d e f V mV Hz...

Page 3: ...is dangerous to ignore the specified warnings Start operations S1 to S4 S1 Install the battery Refer to Section 5 S2 Power on off sequence S3 Example change of function Voltage to Current ON First dis...

Page 4: ...ximise minimise set voltage Menu sequence Task Settings CH1 CH2 Channel Settings Function Units Utility Caption Select Function None Current Voltage A channel can only have one function at a time TAP...

Page 5: ...ic or hydraulic pressures you can have up to two interchangeable pressure transducers at a time Pressure modules PM 620 Optional item These modules attach to the pressure module carrier MC 620 or to a...

Page 6: ...tween the terminals and the ground earth To prevent electrical shocks use only the GE specified AC probe Part IO620 AC to measure AC voltages maximum 300 Vac This instrument uses a Lithium Polymer Li...

Page 7: ...figures on the front cover A2 B1 4 1 Key to figure A2 DPI 620 calibrator Complies with European Union directives Warning refer to the manual Read the manual USB ports Type A Mini type B connector Grou...

Page 8: ...he display is set by the number of task selections and external modules you are working with maximum 6 12 Tap this button to set up the Task set up the instrument Configure and to access Help Refer to...

Page 9: ...onnections and then tighten to the applicable torque Maximum torque 1 8 NPT 35 Nm 26 lbf ft G1 8 25 Nm 18 4 lbf ft A3 Step Procedure 1 When the power is off loosen the five screws a and remove the cov...

Page 10: ...ay with touch screen Operating temperature 10 to 50 C 14 to 122 F Storage temperature 20 to 70 C 4 to 158 F Ingress Protection IP65 DPI 620 calibrator only Humidity 0 to 90 relative humidity RH non co...

Page 11: ...Customer service Visit our web site www gesensing com...

Reviews: