6
Parts
K0454 Issue 1 - [EN] English
4.2 Key to figure B1 (MC 620 module carrier/PM 620 module) - Optional item
A2
5.
USB type A connector for connections to external peripherals
(USB flash memory or optional external modules).
6.
USB mini-type B connector for communication with a
computer.
7.
+5V DC power input socket. This supply also charges the
battery.
8.
Sealed speaker unit.
9.
Liquid crystal display (LCD): Colour display with touch-screen.
To make a selection, lightly tap on the applicable display area.
a.
Battery indicator
b.
Date and time
10.
CH1: Window for the channel 1 settings and values.
c.
Measure or source
indication
d.
Function
e.
Full scale (FS) range
f.
Function units
11.
Other windows: The number of windows you see on the display
is set by the number of task selections and external modules
you are working with (maximum: 6).
12.
Tap this button to set up the
Task
, set up the instrument
(
Configure
) and to access Help (?). Refer to “Quick
Reference”.
Tap this button to maximise each of the available windows in
sequence. Refer to “Quick Reference”.
Pause
(
II
) or
Play
(
X
)
:
Tap (
II
) to hold (freeze) all the data on
the display. To release the display and continue, tap (
X
).
B1
1.
Pressure connection (G1/8 or 1/8NPT) to attach external
pressure equipment.
2.
Pressure and electrical connections for a pressure module
(PM 620). These are self-seal pressure connections.
3.
Two screws to attach the calibrator (DPI 620).
4.
Electrical connections for the calibrator (DPI 620).
5.
Pressure module (PM 620) with a pressure connection,
reference port (a) and a label. The label includes:
Pressure range
. Example: 20 bar g (g: gauge; a: absolute);
serial number
(S/N);
manufacturer
: name, address, website