K114 Issue No. 9
2 - 6
Druck ADTS 405 User Manual
Note:
The unit will not be damaged if AC power and DC power are connected at the same time.
Flight line units
Make sure that the power supply is off before connecting the power cable and hand
terminal cable.
Note:
The flight line version power cable supplies both the electronics and pump racks.
If required, connect the hand terminal to the connector on the front panel using either
the 2 metre or 18 metre cable.
Note:
Connecting the hand terminal disables the front panel key-pad.
The two fuses, located in the holders and mounted on the front panel, protect the
Electronics Rack and the Pump Rack. The fuses are connected in the live supply circuit
and are rated at:
5A anti-surge HBC 250V
An external earth (ground) cable may be connected to the stud on the front panel of
the pump rack to ensure integrity of the earth (ground) connection.
DC Power Option
The DC power option (nominal 28 V DC), can be an alternative power supply using an
additional connector. This is located on the rear of the rack (ADTS 405) or on the pump
front panel (ADTS 405F) both have the same connector details.
Pin
European
U.S.
Function
Colour
Color
1
Grey
Gray
+28V
2
White
White
0V Return
Protective Earth
Protective Ground
Protective Earth
(Ground)
(Ground)
Summary of Contents for Druck ADTS 405
Page 2: ... General Electric Company All rights reserved ...
Page 4: ...K114 Issue No 9 ...
Page 32: ...K114 Issue No 9 2 10 Druck ADTS 405 User Manual F IGURE 2 4 ADTS 405 G ENERAL VIEW ...
Page 33: ...K114 Issue No 9 Installation 2 11 2 F IGURE 2 5 ADTS 405 R EAR P ANEL VIEW ...
Page 41: ...K114 Issue No 9 Druck ADTS 405 User Manual 3 7 3 Control Mode ...
Page 43: ...K114 Issue No 9 Druck ADTS 405 User Manual 3 9 3 Go to Ground ...
Page 50: ...K114 Issue No 9 3 16 Operation FIGURE 3 6 TRIPLE DISPLAY FIGURE 3 5 DUAL DISPLAY ...
Page 72: ...K114 Issue No 9 4 6 Maintenance INTENTIONALLY LEFT BLANK ...
Page 92: ...K114 Issue No 9 5 20 Testing and Fault Finding INTENTIONALLY LEFT BLANK ...