K114 Issue No. 9
Installation
2 - 5
2
2.3 Electrical Connection
WARNINGS: 1
VOLTAGES IN EXCESS OF 30 VOLTS (RMS) AC OR 50VOLTS DC,
IN CERTAIN CIRCUMSTANCES, CAN BE LETHAL. CARE MUST BE TAKEN
WHEN WORKING ON LIVE, EXPOSED CONDUCTORS.
2
DO NOT DISCONNECT THE HAND TERMINAL WHEN ENERGIZED IN THE
HAZARDOUS AREA.
Power Supply Connection
The unit must be connected to the correct electrical power supply as stated, adjacent
to the power connector.
C
AUTIONS
:
1
T
HE
SUPPLY
MUST
PROVIDE
CONNECTION
TO
A
PROTECTIVE
GROUND
TERMINAL
. T
HE
UNIT
MUST
,
AT
ALL
TIMES
,
BE
CONNECTED
TO
THE
SUPPLY
EARTH
(
GROUND
).
2
T
HE
POWER
SUPPLY
CABLE
AND
CONNECTOR
MUST
BE
CORRECTLY
RATED
FOR
THE
POWER
SUPPLY
.
Note:
The ADTS 405 is normally supplied with an approved power supply cable
for use in the country of delivery. This can limit the maximum supply
voltage that can be safely used.
e.g.
a NEMA 5-15P terminated cable, for use in the U.S.A., is approved for a
maximum of 125 V ac; it must be replaced for a higher supply voltage.
ADTS 405 rack mounted units
Make sure that the power supply is off before connecting the power cable .
If required, connect the hand terminal to the connector using either the 2 metre or 18
metre cable.
Note:
Connecting the hand terminal disables the front panel key-pad.
Fit the expansion port plug (item 13, Table 2-1) to the rear panel expansion port.
Note:
For units used with the Druck PV 103 Pump Unit connect the expansion cable, supplied
with the PV 103, to the expansion port.
Pin
European
U.S.
Function
Colour
Color
1
Brown
Black
Live
4
Blue
White
Neutral
Centre
Green/Yellow
Green
Protective Earth
(Ground)
Summary of Contents for Druck ADTS 405
Page 2: ... General Electric Company All rights reserved ...
Page 4: ...K114 Issue No 9 ...
Page 32: ...K114 Issue No 9 2 10 Druck ADTS 405 User Manual F IGURE 2 4 ADTS 405 G ENERAL VIEW ...
Page 33: ...K114 Issue No 9 Installation 2 11 2 F IGURE 2 5 ADTS 405 R EAR P ANEL VIEW ...
Page 41: ...K114 Issue No 9 Druck ADTS 405 User Manual 3 7 3 Control Mode ...
Page 43: ...K114 Issue No 9 Druck ADTS 405 User Manual 3 9 3 Go to Ground ...
Page 50: ...K114 Issue No 9 3 16 Operation FIGURE 3 6 TRIPLE DISPLAY FIGURE 3 5 DUAL DISPLAY ...
Page 72: ...K114 Issue No 9 4 6 Maintenance INTENTIONALLY LEFT BLANK ...
Page 92: ...K114 Issue No 9 5 20 Testing and Fault Finding INTENTIONALLY LEFT BLANK ...