Keypad Display
© 2017 General Electric Company – All Rights Reserved.
ADTS 405 R/F Mk2 Instruction Manual-English | 79
Various commands prompts the operator by displaying messages on screen such as:
“Press the ETP key”. The “ETP” key is the general key for moving onto the next step.
Note:
The
CLEAR/QUIT
key should only be used to terminate test program
execution.
d.
At the end of each test, the operator has a choice of continuing, exiting test program
execution or going to manual mode. In manual mode, any of the normal keypad
functions may be used.
e.
When the manual testing is complete, press
EXECUTE TEST PROGRAM
and the test
program continues from where it left off. This feature may be disabled by including
the KEYLOCK command in the test program. While in manual mode the display
prompts “TPM”.
f.
When a “go” command is executed, “GO” flashes in the lower right hand corner of the
screen. At this point the user should record any instrument readings or displayed leak
rates, as required, before continuing by pressing EXECUTE TEST PROGRAM. F4 may be
used to continue at a “go” command as an alternative to pressing
EXECUTE TEST
PROGRAM
.
7.2.18.2 Loading Program Files
New program files may be prepared on a PC and transferred to the ADTS using the Ethernet
port access procedure described in Section 4.13.3, “Access ADTS Files,” on page 43.
7.2.18.3 Deleting Program Files
If a test program within an ADTS 405 Mk2 is no-longer required, then it may be deleted. To
delete a test program, proceed as follows:
a.
Enter configuration (CONFIG) by pressing and holding
F1
and pressing
SETUP
.
b.
At the first menu level of configuration, press the
EXECUTE TEST PROGRAM
key,
then F2. The display then shows the file name and title of the first file in the list. Use
[NEXT] and [PREV] to step through the files until the display shows the file to be
deleted.
c.
Select [DEL] to delete the file, then press
CLEAR/QUIT
to quit from configuration.
d.
Select [YES] to confirm quit.
Note:
If the deleted program is marked as “AUTO RUN” then an additional screen
asks to confirm if the auto run status should be removed before deletion.
7.2.18.4 Auto Running a Program
To run a program each time the ADTS 405 Mk2 is switched on, the “Auto run” feature should
be enabled. The enabled program starts, from the first line, after successful completion of the
ADTS power-up sequence. To setup the auto run feature, proceed as follows:
a.
Enter configuration (CONFIG) by pressing and holding
F1
and pressing
SETUP
.
b.
At the first menu level of configuration, press the
EXECUTE TEST PROGRAM
key,
then F1. The display shows the name of the current file to “Auto run” or “*NONE*” if
auto run is not selected. To change this selection use the [NEXT] and [PREV] keys and
when the correct file is displayed press [SEL].