8. Specification
Table 8-1: Rack-mount ADTS 405R Mk2 Specification
Parameter
Value
Size
Standard 19" rack front panel:
6U high (10.50”)
Depth behind front panel:
255 mm (10.04”)
Rear panel connections available:
Yes
Weight
a
a. Nominal value. Actual value may vary due to configured optional extras.
12 kg (26.5 lb)
EMC
EN 61326-1
Electrical Safety
EN 61010-1
AC Power Supply
Single phase AC in the range
(automatic range switching):
100/120/230 V, 50/60 Hz
115 V, 400 Hz
Power:
400 VA
Installation Category:
II
Fuse:
T5AH250V
DC Power Supply
Supply range:
16 to 30 V
Power:
400 VA
Fuse:
Not user serviceable.
Operating Environment
Indoor use.
Operating Altitude
Up to 4,570 metres (15,000 ft)
Operating Temperature
-20°C to 50°C (-4°F to 122°F)
Storage and Shipping
Temperature
-51°C to 70°C (-60°F to 158°F)
Ingress Protection
IP23
Operating Humidity
0 to 95% RH (non-condensing)
Pollution Degree
3
Pressure Safety
Pressure Equipment Directive Class: Sound Engineering Practice (SEP)
Shock and Vibration
ETSI EN 300 019-2-5 Class 5.1
Ground Vehicle Installations. Test
method: EN 60068-2-64 Random
Vibration.
Pressure Fittings
Ps:
MS33656-6
Pt:
MS33656-4
Pressure Supply:
MS33656-4
Vacuum Supply:
MS33656-6