Chapter 2. Error Codes and Screen Messages
22
DigitalFlow™ GS868 Service Manual
2.18 Screen Messages
A variety of messages may appear on the display screen during the performance of a task. Since the error codes have
already been discussed in this chapter and the locator bar messages are discussed in detail in Chapter 3,
Operation
, of
the
Startup Guide
, they will not be repeated here. All other messages are listed in Table 2-1 below.
Table 2: Screen Messages
Message
Meaning
Power Up Messages
Backup Battery FAIL
The backup battery that powers the non-volatile RAM has failed.
Contact the factory.
Backup Battery OK
The backup battery that powers the non-volatile RAM has passed.
DSP Processor OK
The DSP processor has passed.
DSP RAM Failed
The DSP RAM has failed. Contact the factory.
DSP RAM OK
The DSP RAM has passed.
EPROM sum =
XXXX
Record the EPROM sum at initial power up and periodically thereafter.
FRIGID_INIT
Executed
The NVR was automatically initialized due to a memory fault.
Contact the factory.
NVR FAIL
The non-volatile RAM has failed. Contact the factory.
NVR OK
The non-volatile RAM has passed.
Measurement Mode Messages
All Logs Cleared!
hit key
This message may appear during one of the following tasks:
1. clearing a log - there are no more logs to clear
2. logging data - user has hit
[ENT]
instead of selecting an old log or entering
a new name
3. printing a log - there are no logs to print
4. displaying a log - there are no logs to display
Hit any key to resume taking measurements.
All Sites Cleared!
hit key
This message may appear during one of the following tasks:
1. saving a new site - a new site name was not entered or overwrite existing
site was not selected
2. recalling a site - there are no site files to recall
3. clearing a site - there are no sites to delete
4. printing a site - there are no sites to print
5. displaying a site - there are no sites to display
Hit any key to resume taking measurements.
Do you want to
SAVE?
This message appears upon exiting the
User Program
, if the new site data
has not been saved. Failure to save will result in loss of the new data next
time the site data is changed or recalled.
Duplicate name. Enter
another.
The site file or log name is already in use. Enter a different name.
Summary of Contents for DigitalFlow GS868
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 10: ...Contents viii DigitalFlow GS868 Service Manual...
Page 26: ...Chapter 1 Calibration DigitalFlow GS868 Service Manual 16...
Page 40: ...Chapter 3 Diagnostics 30 DigitalFlow GS868 Service Manual no content intended for this page...
Page 68: ...Warranty 58 DigitalFlow GS868 Service Manual no content intended for this page...
Page 72: ......
Page 73: ......