DigitalFlow™ GS868 Service Manual
5
Chapter 1. Calibration
1.4.3 Testing the Analog Output Linearity (cont.)
If the linearity test readings are not within 5
A of the values listed in Table 1 above, check the accuracy and wiring of
the ammeter. Then, repeat the low and high end calibrations. If the analog output still does not pass the linearity test,
contact the factory for assistance.
Procedure Options
You have completed calibration of the analog output. Do one of the following:
•
To calibrate or test another output, press
[F4] (EXIT)
and return to Step 3 in
Accessing the Calibration Menu
on
•
To calibrate/test additional slots, press
[F4] (EXIT)
twice and proceed to the appropriate section.
•
To leave the Calibration Menu, press
[F4] (EXIT)
three times.
Table 1: Expected Ammeter Readings
% Full Scale
4-20 mA Scale*
0-20 mA Scale*
0
4.000
0.000
10
5.600
2.000
20
7.200
4.000
30
8.800
6.000
40
10.400
8.000
50
12.000
10.000
60
13.600
12.000
70
15.200
14.000
80
16.800
16.000
90
18.400
18.000
100
20.000
20.000
* All ammeter readings should be ±0.005 mA
Summary of Contents for DigitalFlow GS868
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 10: ...Contents viii DigitalFlow GS868 Service Manual...
Page 26: ...Chapter 1 Calibration DigitalFlow GS868 Service Manual 16...
Page 40: ...Chapter 3 Diagnostics 30 DigitalFlow GS868 Service Manual no content intended for this page...
Page 68: ...Warranty 58 DigitalFlow GS868 Service Manual no content intended for this page...
Page 72: ......
Page 73: ......