Chapter 1. Calibration
1.8 Testing the Totalizer/Frequency Outputs
Totalizer/Frequency outputs may be added to the Model GS868 by installing a
Totalizer/Frequency Option Card
in one
(or more) of the six expansion slots. Each option card includes four outputs, which are designated as A, B, C and D. To
test the outputs, connect a frequency counter to the card’s terminal block as shown in Figure 5 below. Refer to the menu
map in Figure 6 on page 15 as a guide.
Figure 5: Frequency Counter Connections
For this discussion, assume that the option card has been installed in
Slot x
.
1.8.1 Accessing the Calibration Menu
1.
Press the
[CAL]
key to enter the
Calibration Program
.
2.
Press
[Fx]
to calibrate the desired slot. (The option bar will include a slot listing for each installed option card.)
3.
Press
[F1]-[F4]
to select output A, B, C or D, respectively.
1.8.2 Entering Frequency
4.
Enter a new frequency in the range of 1-10,000 Hz and press the
[ENT]
key. Verify that the frequency counter reads
the correct value.
Terminal Block (Option Card)
A
NO
A
COM
B
B
D
D
C
C
Frequency
1
8
Counter
IN
COM
COM
NO
COM
NO
COM
NO
Summary of Contents for DigitalFlow GS868
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 10: ...Contents viii DigitalFlow GS868 Service Manual...
Page 26: ...Chapter 1 Calibration DigitalFlow GS868 Service Manual 16...
Page 40: ...Chapter 3 Diagnostics 30 DigitalFlow GS868 Service Manual no content intended for this page...
Page 68: ...Warranty 58 DigitalFlow GS868 Service Manual no content intended for this page...
Page 72: ......
Page 73: ......