Appendix C. Programming with PanaView
110
DigitalFlow™ GF868 Programming Manual (2-Channel)
C.5.6 Handling Site Files
To access site files, click on the
Site Files
button in the
Properties
window. The
Site File Operations
window (shown in
Figure 31:
Site File Operations
Window
C.5.6a Saving an Existing Site to the Meter
To save an existing site to the meter:
1.
Select the radio button for
Selected
and highlight an existing site in the left pane.
2.
Then click on the
Save Site to Meter
button. A screen opens similar to Figure 32.
Figure 32: Site File Selection
3.
Highlight the desired site and click
Open.
PanaView sends the site to the meter.
Summary of Contents for DigitalFlow GF868
Page 2: ......
Page 4: ...ii no content intended for this page ...
Page 95: ...DigitalFlow GF868 Programming Manual 2 Channel 85 Appendix A Menu Maps Appendix A Menu Maps ...
Page 96: ...86 DigitalFlow GF868 Programming Manual 2 Channel no content intended for this page ...
Page 104: ...Appendix A Menu Maps DigitalFlow GF868 Programming Manual 2 Channel 94 ...
Page 108: ...98 DigitalFlow GF868 Programming Manual 2 Channel no content intended for this page ...
Page 146: ...Warranty 136 DigitalFlow GF868 Programming Manual 2 Channel no content intended for this page ...
Page 150: ......
Page 151: ......