April 2008
1-20
Programming Site Data
Setting Up the Totalizer/
Frequency Outputs
1.
Use the [
←]
and [
→]
and the
[F1]-[F4]
keys to select the desired
slot number.
2.
Press
[F1]-[F4]
to set up outputs A, B, C or D, respectively.
3.
Press
[F1] = OFF
to disable output A and return to the previous
prompt, or press
[F2] = FREQ or [F3] = TTLZR
to designate output A
as a frequency or a totalizer output, respectively.
Note:
The setup of output A is used as an example. Identical
procedures would be used to set up the other outputs.
•
To program a frequency output, go to step 4.
•
To program a totalizer output, go to step 5.
Programming a Frequency
Output
4.
The
FREQ
output,
[F2]
, produces a frequency pulse that is
proportional to the output measurement. Complete the steps below
to program the frequency output.
a.
Use the [
←]
and [
→]
and
[F1]-[F4]
keys to specify the desired
Output Measurement
parameter.
b.
Enter the
Base
value for the low end of the frequency output
range and press
[ENT]
.
c.
Enter the
Full
value for the high end of the frequency output
range. Press
[ENT]
.
d.
Enter a value between 1 and 10,000 for the
Full Scale
Frequency.
Press
[ENT]
.
e.
Use the
[F1]-[F4]
keys to select the desired
Error Option
, the
way the display responds during an error condition. Available
options are shown in Table 1-6 below. Then skip to step 6 on
the next page.
Table 1-6: Error Response Options for Frequency Output
Option
Display Response
HOLD
Holds the last “good” reading.
LOW
Forces the outputs to the low setpoint.
HIGH
Forces the outputs to the high setpoint.
OTHER
Allows user to enter Hertz error level.
Summary of Contents for DigitalFlow GC868
Page 8: ...Chapter 1 ...
Page 61: ...Chapter 2 ...
Page 78: ...Chapter 3 ...
Page 91: ...Chapter 4 ...
Page 105: ...Chapter 5 ...
Page 111: ...Chapter 6 ...
Page 123: ...Appendix A ...
Page 132: ...Appendix B ...
Page 133: ...Data Records Option Cards Installed B 1 Initial Setup Data B 2 ...
Page 136: ...Appendix C ...
Page 156: ...Appendix D ...